H E A D E R

2

0178-1737-000  Rev. A

TABLE OF CONTENTS

User Responsibility 

. . . . . . . . . . . . . . . . . . . . . . . . . .  1

Cross Connection Prevention  . . . . . . . . . . . . . . . . . .  1

Precautions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1

1/Introduction 

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2

2/General Information

2.1

Description  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2 

2.2

Dry Contact Relays  . . . . . . . . . . . . . . . . . . . . . . . .  2

2.3

Specifications  . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2

3/Set-Up and Operation

3.1

Pre-Use Checkout  . . . . . . . . . . . . . . . . . . . . . . . . .  4

3.2

Controls and Indicators  . . . . . . . . . . . . . . . . . . . . .  4

4/Operator Maintenance

4.1

Repair Policy  . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5

4.2

General Maintenance 

. . . . . . . . . . . . . . . . . . . . . .  5

4.3

Testing Switch Gauges 

. . . . . . . . . . . . . . . . . . . . .  6

Warranty  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9

Cross Connection Prevention

After installation, perform a pipeline system check to cer-

tify that there are no cross connections, in accordance

with National Fire Protection Association 99.

User Responsibility

This Product will perform in conformity with the descrip-

tion thereof contained in this operating manual and accom-

panying labels and/or inserts, when assembled, operat-

ed, maintained and repaired in accordance with the

instructions provided. This Product must be checked peri-

odically. A defective Product should not be used. Parts

that are broken, missing, plainly worn, distorted or conta-

minated should be replaced immediately. Should such

repair or replacement be come necessary,

BeaconMedæs recommends that a telephonic or written

request for service advice be made to BeaconMedæs.

This Product or any of its parts should not be repaired

other than in accordance with written instructions provid-

ed by BeaconMedæs and by BeaconMedæs trained per-

sonnel. The product must not be altered without the prior

written approval of BeaconMedæs The user of this

Product shall have the sole responsibility for any mal-

function which results from improper use, faulty mainte-

nance, improper repair, damage, or alteration by anyone

other than BeaconMedæs. 

PRECAUTIONS

Warnings

After installation of an alarm panel, make absolutely cer-

tain that the alarms are correctly connected to their

respective gas lines in accordance with NFPA 99.

Do not use liquid cleaning solutions on this unit. Liquid

solutions may enter the unit and cause system damage or

present an electric shock hazard. See Page 5.

The “Testing the Switch Gauges” procedure requires a

change of switch gauge trigger points during the test.

Readjust the alarm trigger points after completing this

procedure. See Page 6.

Caution

No repair should ever be under taken or attempted by

anyone not having general experience in the repair of

devices of this nature. See Page 5.

Содержание Area Line Pressure Alarm

Страница 1: ... Technical Assistance 0178 1737 000 Rev A Technical Support 1 888 4MEDGAS 463 3427 Customer Service 1 888 4MEDGAS 463 3427 www beaconmedaes com Area Line Pressure Alarm OPERATION MAINTENANCE MANUAL Product No 0321 ...

Страница 2: ...eplacement be come necessary BeaconMedæs recommends that a telephonic or written request for service advice be made to BeaconMedæs This Product or any of its parts should not be repaired other than in accordance with written instructions provid ed by BeaconMedæs and by BeaconMedæs trained per sonnel The product must not be altered without the prior written approval of BeaconMedæs The user of this ...

Страница 3: ... is made at the circuit board One relay indicates normal conditions the other six will indicate alarm condi tions These relays can operate as normally open or normally closed switches depending on how the connections are made Refer to the installation instructions or service manual for the wiring diagram 2 3 Specifications Gases Monitored From one to six of the following gases can be monitored O2 ...

Страница 4: ...L The gauge pointer moves in response to change in gas line pressure Under normal pressure conditions the pointer will not be in con tact with the gauge alarm contacts Switch Gauge Operation ABNORMAL As gas line pressure goes to an abnormal high or low level the pointer will move accordingly Once the pointer touches one of the contacts the normal 5V signal will be inter rupted and an alarm conditi...

Страница 5: ...urs If multiple abnormal conditions exist the alarm will continue to sound until all have been corrected Test Press this button to test the electronics of the panel When this button is pressed the NORMAL and all the ABNORMAL lights should glow The alarm tone should sound Normal This green light glows when all of the gas pres sures and vacuum levels are within preset limits It turns off when an abn...

Страница 6: ...aconMedæs Then test the unit to ascertain that it complies with the manufacturer s published specifications Contact BeaconMedæs Technical Support for service assistance In all cases other than where BeaconMedæs s warranty is applicable repairs will be made at BeaconMedæs s current list price for the replace ment part s plus a reasonable labor charge 4 2 General Maintenance The alarm panel is desig...

Страница 7: ...e should be long enough so that the ends of the tape clear the front panel Figure 8 4 Mark the exact location of the contact assembly s brass pins on the transparent tape Figure 8B 5 Using the 1 16 hex wrench individually turn each low left side contact so it just touches the gauge needle Figure 8C 6 The alarm must sound on contact If no alarm sounds the panel needs repair Figure 8D Note If remote...

Страница 8: ... 1737 000 Rev A 4 OPERATOR MAINTENANCE Figure 7 Gauge Location Note Each line pressure or vacuum gauge has a corre sponding switch gauge Figure 8 Pressure Switch Testing Note For clarity shorting bars are not shown ...

Страница 9: ...nable satisfaction that such equipment has not been damaged in shipment or installa tion improperly installed operated out side of any published operating limits including but not limited to temperature pressure humidity or ventilation improperly or inadequately maintained field modi fied in any way improperly repaired or in any other way improperly applied or used All claims against this warranty...

Страница 10: ... 2004 BeaconMedæs All rights reserved Part No 0178 1737 000 Rev A ...

Отзывы: