1/INTRODUCTION

2/GENERAL INFORMATION

This manual gives the recom-

mended operating instructions

for the BeaconMedæs Area

Line Pressure Alarm Panel.

Every procedure listed is per-

formable by the operator of this

equipment. For service informa-

tion, refer to the BeaconMedæs

Area Line Pressure Alarm Panel

Service Manual.

Figure 1

Area Alarm Circuit

2.1 Description

The area line pressure alarm panel is a wall mounted unit

that monitors the gas pressure (or vacuum) in up to six

separate lines and sounds an alarm if preset limits are

violated (Figure 1).

One panel holds up to six gauges. Each gauge is labeled

with a color-coded label for the gas it monitors. In the

event of an alarm, a red indicator (labeled ABNORMAL)

glows next to the gauge for the affected gas (Figure 2).

Each switch gauge is wired in a closed loop: if a wire

breaks, or is disconnected from the circuit board or

switch, the alarm panel sounds an alarm (Figure 3).

The panel has a test switch that exercises the electronics

assembly. Remove the louvered panels to test the switch

gauges. This means that the unit can be tested for func-

tionality without major disassembly or special test equip-

ment.

Use only approved service procedures and genuine

replace ment parts for repairs. Complete service informa-

tion and a list of available replacement parts are provided

in the Service Manual.

2.2 Dry Contact Relays

The alarm panel contains seven dry contact relays for

interface to remote equipment. The connection to these

relays is made at the circuit board. One relay indicates

normal conditions; the other six will indicate alarm condi-

tions.

These relays can operate as normally open or normally

closed switches depending on how the connections are

made. Refer to the installation instructions or service

manual for the wiring diagram.

2.3 Specifications

Gases Monitored: From one to six of the following gases

can be monitored. O

2

, N

2

O, N

2

, AIR, CO

2

, VAC, EVAC.

Dimensions

Alarm Panel:

14 in. x 9 3/4 in.

(35.6cm x 24.6cm)

Enclosure (rough-in box):

12 5/8 in. x 8 3/8 in. x 3 5/8 in.

(32.5cm x 21.3cm x 9.5 cm)

Color Coding 

Standard, USA color coded labels for all readouts.

Optional Canadian and ISO color coded labels available.

Power Requirements 

Standard 120 V~, 60 Hz, 25 watts (max.) 

Export Only 220 V~, 50 Hz, 25 watts (max.)

Note: 

The 120 V~ model is listed by Underwriters

Laboratories.  The 220 V~ model has not been investi-

gated.

Leakage Current

Less than 100 uA

Circuit Type 

Closed loop-alarm on ground contact or open circuit

Control Unit

Microprocessor-Motorola 6805 Series

Gauge Accuracy:

Gas

Indicating Gauge

Switch Gauge

Nominal 50 psi

±2 psi (25-75 psi)

±2 psi (33-66 psi)

Medical Gases

±3 psi elsewhere

±3 psi elsewhere

Nominal 180 psi

±6 psi (75-225 psi)

±6 psi (100-200 psi)

Gas

±9 psi elsewhere

±9 psi elsewhere

Vacuum Switch

±.6 in. Hg 

(7.5-22.5 in. Hg)

±.6 in. Hg (10-20 in. Hg)

±.9 in. Hg elsewhere

±.9 in. Hg elsewhere

Gauge Overpressure Limit

Indicating Gauge

Switch Gauge

Oxygen

110 psig

150 psig

Nitrous Oxide

110 psig

150 psig

Air

110 psig

150 psig

Carbon Dioxide

110 psig

150 psig

Nitrogen

330 psig

400 psig

Vacuum

(Do not pressurize)

45 psig

Note: 

If line pressure should fall or rise to

set alarm limits the gauge needle will

interrupt current flow and cause an alarm

condition.

Содержание Area Line Pressure Alarm

Страница 1: ... Technical Assistance 0178 1737 000 Rev A Technical Support 1 888 4MEDGAS 463 3427 Customer Service 1 888 4MEDGAS 463 3427 www beaconmedaes com Area Line Pressure Alarm OPERATION MAINTENANCE MANUAL Product No 0321 ...

Страница 2: ...eplacement be come necessary BeaconMedæs recommends that a telephonic or written request for service advice be made to BeaconMedæs This Product or any of its parts should not be repaired other than in accordance with written instructions provid ed by BeaconMedæs and by BeaconMedæs trained per sonnel The product must not be altered without the prior written approval of BeaconMedæs The user of this ...

Страница 3: ... is made at the circuit board One relay indicates normal conditions the other six will indicate alarm condi tions These relays can operate as normally open or normally closed switches depending on how the connections are made Refer to the installation instructions or service manual for the wiring diagram 2 3 Specifications Gases Monitored From one to six of the following gases can be monitored O2 ...

Страница 4: ...L The gauge pointer moves in response to change in gas line pressure Under normal pressure conditions the pointer will not be in con tact with the gauge alarm contacts Switch Gauge Operation ABNORMAL As gas line pressure goes to an abnormal high or low level the pointer will move accordingly Once the pointer touches one of the contacts the normal 5V signal will be inter rupted and an alarm conditi...

Страница 5: ...urs If multiple abnormal conditions exist the alarm will continue to sound until all have been corrected Test Press this button to test the electronics of the panel When this button is pressed the NORMAL and all the ABNORMAL lights should glow The alarm tone should sound Normal This green light glows when all of the gas pres sures and vacuum levels are within preset limits It turns off when an abn...

Страница 6: ...aconMedæs Then test the unit to ascertain that it complies with the manufacturer s published specifications Contact BeaconMedæs Technical Support for service assistance In all cases other than where BeaconMedæs s warranty is applicable repairs will be made at BeaconMedæs s current list price for the replace ment part s plus a reasonable labor charge 4 2 General Maintenance The alarm panel is desig...

Страница 7: ...e should be long enough so that the ends of the tape clear the front panel Figure 8 4 Mark the exact location of the contact assembly s brass pins on the transparent tape Figure 8B 5 Using the 1 16 hex wrench individually turn each low left side contact so it just touches the gauge needle Figure 8C 6 The alarm must sound on contact If no alarm sounds the panel needs repair Figure 8D Note If remote...

Страница 8: ... 1737 000 Rev A 4 OPERATOR MAINTENANCE Figure 7 Gauge Location Note Each line pressure or vacuum gauge has a corre sponding switch gauge Figure 8 Pressure Switch Testing Note For clarity shorting bars are not shown ...

Страница 9: ...nable satisfaction that such equipment has not been damaged in shipment or installa tion improperly installed operated out side of any published operating limits including but not limited to temperature pressure humidity or ventilation improperly or inadequately maintained field modi fied in any way improperly repaired or in any other way improperly applied or used All claims against this warranty...

Страница 10: ... 2004 BeaconMedæs All rights reserved Part No 0178 1737 000 Rev A ...

Отзывы: