5. In installations with more than one vacuum pump connected to a common net, manual valves must be

installed to isolate each pump. Non-return valves (check valves) must not be relied upon for isolating the
pumps.

6. Never remove or tamper with the safety devices.

Also consult following safety precautions: 

Safety precautions during operation

 and 

Safety

precautions during Maintenance or Repair

.

These precautions apply to vacuum pumps.
For precautions applying to the connected equipment consult the relevant instruction book.
Some precautions are general and cover several machine types and equipment; hence some
statements may not apply to your device.

1.3

Safety precautions during operation

All responsibility for any damage or injury resulting from neglecting these precautions, or non-
observance of the normal caution and care required for installation, operation, maintenance and
repair, even if not expressly stated, will be disclaimed by the manufacturer.

General precautions

1. The operator must employ safe working practices and observe all related local work safety requirements

and regulations.

2. If any of the following statements does not comply with local legislation, the stricter of the two shall apply.
3. Installation, operation, maintenance and repair work must only be performed by authorised, trained,

specialised personnel.

4. Before carrying out any maintenance, repair work, adjustment or any other non-routine checks, stop the

device. In addition, the power isolating switch must be opened and locked.

Precautions during operation

1. Persons switching on remotely controlled machines shall take adequate precautions to ensure that there

is no one checking or working on the machine. To this end, a suitable notice shall be affixed to the remote
start equipment.

2. Never operate the device in the presence of flammable or toxic fumes, vapours or particles.
3. Never operate the machine below or in excess of its limit ratings.
4. Wear ear protectors if applicable. People staying in environments or rooms where the sound pressure level

reaches or exceeds 90 dB(A) shall wear ear protectors.

5. Periodically check that:

• All guards and fasteners are in place and tight
• All hoses and/or pipes are in good condition, secure and not rubbing
• There are no leaks
• All electrical leads are secure and in good order

6. Never remove or tamper with the safety devices.

Instruction book

5

Содержание AGS 1560

Страница 1: ...Instruction Book Operating Manual AGS Anaesthetic Gas Scavenging Disposal System Part Number 2920 7087 61 Revision Date 1 1 2011...

Страница 2: ......

Страница 3: ...ying of the contents or any part thereof is prohibited This applies in particular to trademarks model denominations part numbers and drawings This instruction book is valid for CE as well as non CE la...

Страница 4: ...2 EXHAUSTER UNITS 10 2 3 ELECTRIC MOTORS 12 2 4 CONTROL PANEL 12 2 5 REMOTE START SWITCH PANELS 18 2 6 DRAIN AND VALVE ASSEMBLY 18 2 7 EXHAUST SILENCER 18 2 8 FLOW REGULATING VALVE 19 2 9 TERMINAL UN...

Страница 5: ...8 Technical data 38 8 1 REFERENCE CONDITIONS AND LIMITATIONS 38 8 2 ELECTRIC DATA 39 8 3 PERFORMANCE DATA 40 8 4 PRESSURE SWITCH SETTINGS 42 9 Declaration of conformity 43 Instruction book 3...

Страница 6: ...addition the power isolating switch must be opened and locked Precautions during installation 1 Place the device where the ambient air is as cool and clean as possible within the limitations for opera...

Страница 7: ...d regulations 2 If any of the following statements does not comply with local legislation the stricter of the two shall apply 3 Installation operation maintenance and repair work must only be performe...

Страница 8: ...cks stop the device In addition the power isolating switch must be opened and locked Precautions during maintenance or repair 1 Use only the correct tools for maintenance and repair work 2 Use only ge...

Страница 9: ...ns applying to the connected equipment consult the relevant instruction book Some precautions are general and cover several machine types and equipment hence some statements may not apply to your devi...

Страница 10: ...ressure drop min for low flow systems Waste anaesthetic gas diluted by room air within the attached receiving system is transmitted by vacuum which is generated by an exhauster unit Vacuum levels with...

Страница 11: ...Simplified schematic of a AGS Simplex installation Instruction book 9...

Страница 12: ...gas is discharged safely to atmosphere Eventually an optional exhaust silencer can be fitted to the discharge opening The electric motor and impeller assemblies are directly coupled and the motor bea...

Страница 13: ...AGS Simplex AGS Duplex Instruction book 11...

Страница 14: ...l circuit electrical power supply at 24 V AC A green POWER ON indicator H1 is fitted to the starter isolator panel and is illuminated whenever power is available to the 24 V control and indication cir...

Страница 15: ...Control panel AGS Simplex S1 Main switch H1 Power ON lamp H2 Motor RUNNING lamp S10 HAND OFF AUTO switch Instruction book 13...

Страница 16: ...Service diagram AGS Simplex 1 ph typical Instruction book 14...

Страница 17: ...Fuses F21 Overload protection H1 Lamp voltage on H2 Lamp motor running K21 Main contactor RB1 Rectifier bridge K26 K27 Auxiliary relay S1 Main switch S10 HAND OFF AUTO switch T1 Control transformer 1X...

Страница 18: ...K10 BMS plant emergency contact option not shown 1X3 Terminal M1 Motor PS01 Electronic pressure switch Duplex installation Control panel AGS Duplex Instruction book 16...

Страница 19: ...usly described but incorporates additional features An additional switch S enables selection of a Duty pump and sets a particular exhauster unit to operate the system whilst the second exhauster unit...

Страница 20: ...s with pressure switch the indicator light is also extinguished if the vacuum levels fall below a pre determined value All remote switches must be selected OFF before the exhauster unit will shut down...

Страница 21: ...a function of system design to achieve the required performance Each valve by controlling the vacuum level ensures that the flow rate at the terminal unit is maintained within the design limits regard...

Страница 22: ...acuum switch PN7009 For general information on the electronic vacuum switch type used is PN7009 please consult also the dedicated manual 2920 7089 50 on the CD delivered with the AGS unit For the fact...

Страница 23: ...the maximum flow rate to approximately 150 litres min This provides an ultimate safety device to protect the patient and downstream equipment 2 10 Duplex installation Duplex installations consist of...

Страница 24: ...ox are compatible with the local power supply Before electrically connecting the AGS plant ensure that the power supply is off and correctly isolated The electrical power supplies to the AGS plant mus...

Страница 25: ...be glued on each drain flask 4 see section Dimension drawings at installation Use Loctite 577 on conical threads prior to assembly Cleaning of the piping network required prior to installation of the...

Страница 26: ...Warning label biological hazard 3 2 Dimension drawings AGS Simplex 1 Flexible hose 4 Drain flask Instruction book 24...

Страница 27: ...ng valve may be swapped with the inlet suction side see Installation proposal drawings Alt 1 and Alt 2 section Introduction installation Connect the distribution pipeline installation to the suction i...

Страница 28: ...zed cable 4 Check operation Check operation of exhauster unit in accordance with the commissioning procedure 5 Remote start switches Position remote switches in appropriate places and wire them in acc...

Страница 29: ...m the distribution system if previously fitted Use one open pipeline to pressure test the system and blank off the remainder 6 Disconnect the inlet pipeline to the exhauster unit and blank off the ope...

Страница 30: ...vel achieves the preset value of the plant failure pressure switch 8 Ensure that the exhauster unit runs smoothly without any unusual noise and switch it OFF 9 Remove the terminal block cover on the e...

Страница 31: ...ow rate is 130 litres min 9 Repeat the process detailed is section 5 to 8 above for each terminal unit in succession back to the one closest to the plant ensuring the terminal unit flow does not excee...

Страница 32: ...pen the three phase MCB Miniature Circuit Breaker in each panel Close both panels and switch isolators back to ON Switch a remote switch to ON and check that after 15 seconds the red PLANT EMERGENCY i...

Страница 33: ...ructions Remote switch control Control panel AGS Simplex Control panel AGS Duplex Ensure that the electrical power supply to the starter isolator panel is ON and select the isolating switch to ON Inst...

Страница 34: ...switches are selected OFF Continuous operation With the electrical power supply ON and the isolating switch selected ON the exhauster unit operates when the HAND OFF AUTO switch is selected to HAND T...

Страница 35: ...air cooling inlet and exhaust are free from obstruction Run the exhauster unit and ensure that its operation is smooth and free from any unusual noise levels Drain and clean the drain assembly as nec...

Страница 36: ...unit Clean the mesh disc and replace filters as necessary 2 Flow regulating valve Clean the mesh disk Grease the shaft with Klubersynth UH1 151 If the flow regulating valve is equipped with an air fi...

Страница 37: ...uring commissioning and system performance checks Routine maintenance consists of a visual inspection for security and damage If valve performance is suspect it may be stripped inspected for damage an...

Страница 38: ...d tighten pipe connections System flow rate too low and cannot be reset Exhauster unit filters blocked Remove and inspect the exhauster unit filters for obstruction Flow regulating valve not seating I...

Страница 39: ...Condition Cause Remedy Unserviceable electrical logic circuits within control panel Check the control panel for fault or replace the control panel Instruction book 37...

Страница 40: ...atmospheric pressure mbar a 800 Minimum atmospheric pressure psi a 11 61 Maximum ambient temperature C 40 Maximum ambient temperature F 104 Minimum ambient temperature C 1 Minimum ambient temperature...

Страница 41: ...30 1 400V N 50 3 ph 3 00 7 2 7 9 1 5 16 AGS 3370 S SV7 3 30 1 380V N 60 3 ph 3 60 7 2 7 9 1 5 16 AGS Duplex Cubicle Type Pump type Voltage V Freq Hz Phase Max power kW Nom current A Over load setting...

Страница 42: ...ired at reference conditions l min 533 533 1560 2900 Motor power data based on shaft input at reference conditions kW 0 38 0 38 1 10 1 49 Motor power data based on shaft input at reference conditions...

Страница 43: ...ditions hp 0 65 1 87 2 65 Mean sound pressure level at reference conditions dB A 64 69 70 Design data AGS 520 S AGS 650 S 1 ph Simplex AGS 520 S AGS 650 S 3 ph Simplex AGS 1560 S AGS 2210 S 3 ph Simpl...

Страница 44: ...put 1 Fno SP2 Switch on point 2 70 rP2 Switch off point 2 980 OU2 Switching function for output 2 Fno dS1 Switch on delay for output 1 15 dr1 Switch off delay for output 1 dS2 Switch on delay for outp...

Страница 45: ...9 Declaration of conformity Typical example of a Declaration of Conformity document Instruction book 43...

Страница 46: ......

Страница 47: ......

Страница 48: ......

Отзывы: