AGS Duplex
1
Flexible hose
4
non-return valve
5
Drain flask
3.3
Installation procedure
1.
Locate the AGS unit:
Move the frame mounted unit into position and drill the floor to receive anchor nuts. Fit mounting bolt
anchor nuts into the floor. Choose a suitable position where support for the flexible pipe work can be
fitted.
Fit mounting bolts and washers and nuts and fully tighten down.
2.
Connect piping:
•
Clean the piping system before connecting it to the AGS unit.
•
If more convenient, the flow regulating valve may be swapped with the inlet suction
side (see Installation proposal drawings, Alt. 1 and Alt. 2 - section
Introduction
installation
.
Connect the distribution pipeline installation to the suction inlet. Connect the exhaust pipeline installation
to the outlet connection, ensuring that the drain valve assembly is perpendicular to the floor.
The flexible hose will not support the weight of the connecting pipe work. The connecting pipe work must
be secured to give added stability.
The diameter of the outlet pipe must be large enough to avoid pressure buildup. Back pressure has a
negative impact on the performance of the pumps and must be limited as much as possible (max. 10 mbar).
Instruction book
25
Содержание AGS 1560
Страница 2: ......
Страница 11: ...Simplified schematic of a AGS Simplex installation Instruction book 9...
Страница 13: ...AGS Simplex AGS Duplex Instruction book 11...
Страница 16: ...Service diagram AGS Simplex 1 ph typical Instruction book 14...
Страница 45: ...9 Declaration of conformity Typical example of a Declaration of Conformity document Instruction book 43...
Страница 46: ......
Страница 47: ......
Страница 48: ......