Chapter 5: Optimizing cytometer settings
67
2. If you need to select a configuration other than the current
configuration:
a. In the
Configurations
tab, select a configuration.
b. Click
Set Configuration.
c. Click
OK.
d. Verify that the configuration you just set matches your flow
cytometer optics.
3. Click
OK
to close the
Cytometer Configuration
window.
4. Select
File > Exit
to close
CS&T
.
5. Select
Edit > User Preferences
.
6. Click the
General
tab and select the
Load data after recording
checkbox.
See the
BD FACSDiva Software Reference Manual
for more
information about cytometer configurations and user preferences.
Next step
Running a performance check (page 68)
More information
•