User’s Guide
80
6. Verify that the positive populations are on scale.
If a positive population is off scale, lower the PMT voltage for
that parameter until the positive population can be seen
entirely on scale.
7. Unload the multicolor sample.
8. Place a tube containing DI water on the SIP and put the
cytometer on standby.
9. (Optional) Save the application settings by right-clicking
Cytometer settings
in the
Browser
, then selecting
Cytometer
Settings > Save.
10. In the
Save Cytometer Settings
dialog, enter a descriptive name
for the settings.
11. Click
OK
.
The application settings are saved to the catalog.
Next step