Cleaning the fluidics
Introduction
This topic describes how to perform the daily fluidics cleaning.
Overview
Cleaning the fluidics daily prevents the sample injection tube from becoming clogged and removes dyes that
can remain in the tubing.
In addition to daily cleaning, follow this procedure immediately after running viscous samples or nucleic acid
dyes such as Hoechst, DAPI, propidium iodide (PI), acridine orange (AO), or thiazole orange (TO).
Procedure
To clean the fluidics:
1. Press RUN and HIGH on the cytometer fluid control panel.
2. Install a tube containing 3 mL of a 10% bleach solution on the SIP with the support arm to the side (vacuum
on) and let it run for 1 minute.
3. Move the tube support arm under the tube (vacuum off) and allow the cleaning solution to run for 5
minutes with the sample flow rate set to HIGH.
4. Repeat steps 2 and 3 with a 1.5% dilution of BD
®
Detergent Solution Concentrate.
Note:
The BD
®
Detergent Solution Concentrate must be diluted before use. Mix one full 15 mL bottle of
BD
®
Detergent Solution Concentrate into 985 mL of DI water to make 1 L total.
Never mix BD
®
Detergent Solution Concentrate and bleach because they create chlorine gas.
5. Repeat steps 2 and 3 with DI water.
6. Press the STANDBY button on the fluidics control panel.
7. Place a tube containing less than 1 mL of DI water on the SIP.
A tube with less than 1 mL of DI water should remain on the SIP to prevent salt deposits from forming in the
injection tube. This tube also catches back drips from the flow cell.
Do not leave more than 1 mL of water on the SIP. When the instrument is turned off or left in
standby mode, a small amount of fluid will drip back into the sample tube. If there is too much
fluid in the tube, it could overflow and affect the cytometer performance.
Chapter 4 Maintenance
39
Содержание FACSymphony A1 Flow
Страница 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Страница 6: ......
Страница 10: ...This page intentionally left blank ...
Страница 24: ...This page intentionally left blank ...
Страница 36: ...This page intentionally left blank ...
Страница 50: ...This page intentionally left blank ...
Страница 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Страница 88: ...This page intentionally left blank ...
Страница 96: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...