About the examples
The examples in this chapter use a 4-color bead sample with the following fluorochromes:
l
FITC
l
PE
l
PerCP-Cy5.5
l
APC
If you follow this workflow with a different bead sample (or another sample type), your software views, data
plots, and statistics might differ from the example. Additionally, you might need to modify some of the
instructions in the procedure.
The information shown in italics is for example only. You can substitute your own names for folders and
experiments.
Verifying the configuration and user preferences
Introduction
This topic describes how to verify the cytometer configuration and user preferences before you create an
experiment.
To obtain accurate data results, the current cytometer configuration must reflect your
BD FACSymphony™ A1 flow cytometer optics.
Chapter 5 Optimizing cytometer settings
53