BDDF00122 Issue 3

21/36

Alaris™ Gateway Workstation v1.3.x

Operating Precautions

Operating Precautions

Operating Environment

•  Users of the Workstation should read all instructions in this manual before using this medical device .
•  The Workstation is suitable for all establishments, including those directly connected to the public low-

voltage power supply network . The Workstation is designed for use in the hospital environment, not a road 

ambulance or home health care environments .

•  When setting up a Workstation, an assessment of any potential hazards associated with the routing 

of electrical leads and infusion lines should be made . Where appropriate, mitigations identified and 

implemented .

•  The Workstation should only be used with compatible BD products, accessories, infusion bags and 

disposables .

•  While being used for patient therapy, each Workstation should be dedicated to the care of a single patient . 

Where an auxiliary Workstation is used, this should be dedicated to the same patient as the primary 

Workstation to which it is connected .

•  This Workstation is not intended to be used in the presence of a flammable anaesthetic mixture with air or 

oxygen or nitrous oxide .

•  The Workstation does not support antivirus software . It is the Hospital’s responsibility to ensure that the 

network is secure .

•  Special instructions for isolation room:

 

– Use with the nurse-call feature in order to make alarms visible outside of the room

 

– Exclude use of the Alaris GW Volumetric Pumps, which are not fully compatible with the 

Workstation nurse call .

Electromagnetic Compatibility and Interference

M

•  This Workstation is protected against the effects of external interference, including high energy 

radio frequency emissions, magnetic fields and electrostatic discharge (for example, as generated by 

electrosurgical and cauterising equipment, large motors, portable radios, cellular telephones etc .) and has 

been tested to IEC/EN60601-1-2 and ETSI EN 301 489-17 (where applicable) .

•  The Workstation is a CISPR II Group 1 Class A device . When Alaris Infusion Pumps are attached and 

operational, the system becomes a CISPR II Group 1 Class A system .

•  This Workstation is a CISPR II Group 1 Class A equipment and uses RF energy only for its internal function 

in the normal product offering . Therefore, its RF emissions are very low and are not likely to cause any 

interference with the nearby electronic equipment . However, this Workstation emits a certain level of 

electromagnetic radiation which is within the levels specified by IEC/EN60601-2-24, IEC/EN60601-1-2 

and ETSI EN 301 489-17 (where applicable) . If however the Workstation interacts with other equipment, 

measures should be taken to minimise the effects, for instance by repositioning or relocation .

• 

Therapeutic Radiation Equipment:

 Do not use the Workstation in the vicinity of any Therapeutic 

Radiation Equipment . Levels of radiation generated by the radiation therapy equipment such as a 

Linear Accelerator, may severely affect functioning of the Workstation . Please consult manufacturer’s 

recommendations for safe distance and other precautionary requirements . For further information, please 

contact your local BD representative .

MR

• 

Magnetic Resonance Imaging (MRI):

 The Workstation contains ferromagnetic materials which are 

susceptible to interference with magnetic field generated by the MRI devices . Therefore, the Workstation is 

considered not to be MRI compatible . If use of the Workstation within an MRI environment is unavoidable, 

then BD highly recommends securing the Workstation at a safe distance from the magnetic field outside 

the identified ‘Controlled Access Area’ in order to evade any magnetic interference to the Workstation; 

or MRI image distortion . This safe distance should be established in accordance with the manufacturers’ 

recommendations regarding electromagnetic interference (EMI) . For further information, please refer to 

the product technical service manual (TSM) . Alternatively, contact your local BD representative for further 

guidance .

• 

Accessories:

 Do not use any non-recommended accessory with the Workstation . The Workstation is 

tested and compliant with the relevant EMC claims only with the recommended accessories . Use of any 

accessory, transducer or cable other than those specified by BD may result in increased emissions or 

decreased Workstation immunity .

•  The Workstation features an optional radio frequency IEEE 802 .11n Wireless Local Area Network interface 

(Universal Dual Band Wireless Internet Adapter) . When fitted, the Workstation must emit electromagnetic 

energy in order to perform its intended function . Nearby electronic equipment may be affected .

•  In some circumstances the Workstation may be affected by an electrostatic discharge through air at 

levels above 15kV; or by radio frequency radiation above 10V/m . If the Workstation is affected by this 

external interference, it will remain in a safe mode and alerts the user by generating a combination of 

visual indicators and audible tones . Should any encountered alarm condition persist, even after user 

intervention, it is recommended to replace that particular Workstation and quarantine for the attention of 

Qualified Service Personnel .

•  Portable and mobile RF communications equipment can affect other, nearby medical electrical equipment .

Содержание Alaris Gateway Workstation v1.3.x

Страница 1: ...u Alaris Gateway Workstation v1 3 x Directions For Use en...

Страница 2: ...to Infusion Pumps 17 Powering an auxiliary Workstation 17 System Notifications 18 System Fault Indication 19 Nurse Call Interface 19 Pump Alarm Location Beacon where fitted 20 Operating Precautions 2...

Страница 3: ...t Interface where fitted 30 Auxiliary Interface 30 Product Specifications 31 Electrical 31 Battery 31 Environmental 31 Classification 31 Regulatory Compliance 31 Potential Equalisation Conductor 31 Pa...

Страница 4: ...32 SN T 135070253 17 04 2014 This is an entry level Workstation with organization and central AC power supply functionality It does not support PDMS connectivity It may be used as an Auxiliary Worksta...

Страница 5: ...h has an odd number and that the AC power Cable socket is at the front Note Please make sure to refer to the correct Workstation Directions For Use to familiarise with the specific Workstation version...

Страница 6: ...tory alarms and warnings from software versions prior to 3 4 5 3RD EDITION ALARMS Low medium and high priority alarm tones in accordance with IEC 60601 1 8 2012 and IEC 60601 2 24 2012 Note The 3rd Ed...

Страница 7: ...p with Plus Software 9002TIG03 v2 3 x Yes Yes Yes Yes Yes Yes Alaris GP Guardrails Volumetric Pump with Plus Software 9002TIG03 G v2 3 x Yes Yes Yes Yes Yes Yes Alaris VP Plus Guardrails Volumetric Pu...

Страница 8: ...8002MED01 8002TIG01 v4 1 8 Yes Yes Yes Yes Alaris CC Syringe Pump with Plus Software MK4 8003MED01 8003TIG01 v4 1 8 Yes Yes Yes Yes Alaris GH Guardrails Syringe Pump with Plus Software MK4 8002MED01 G...

Страница 9: ...Pump Alaris GW 800 Volumetric Pump Condition Alarm Location Beacon Behaviour Alarm Location Beacon Behaviour End of Infusion KVO enabled Amber Beacon in KVO Phase None for KVO Phase End of Infusion K...

Страница 10: ...0300UNSxx where xx indicates the connectivity option The two or three numbers at the end of the SKU identifies the Workstation configuration see below for all available configurations Available Config...

Страница 11: ...e can be obtained free of charge by contacting your local BD representative An estimated delivery time will be provided when the order is placed Conventions used in this manual BOLD Used for Display n...

Страница 12: ...tion j j Battery Indicator When illuminated the Workstation is operating from internal battery when flashing the battery power is low and auto power down is imminent k k AC Power Indicator When illumi...

Страница 13: ...onnector IP22 Protection against fingers or other object not greater than 80mm in length and 12mm in diameter Protection against water droplets deflected up to 15 from vertical Protected against acces...

Страница 14: ...able Height Pole Slip Catch Pump Alarm Location Beacon Not fitted on 4 MDI Tile Horizontal Module Bag Supports w Alaris Gateway Workstation v1 3 IP22B Covered under Patents U S Pat 6 593 528 EP 150261...

Страница 15: ...ion Configuration 80300UNSxx 235 shown mounted on an Alaris Trolley Infusion Line Tidies Bag Supports Trolley Bag Support Rail Mounting Rail Mounting Rail Mounting Rail Note Qualified Service Personne...

Страница 16: ...ure the device 3 Refit rubber strip Note It is recommended that line tidies be used to organize the infusion lines this could reduce the risk of tangled infusion lines and tripping the user or patient...

Страница 17: ...double fuse holder located on the AC inlet connector Warning When connected to the AC supply a three wire Live Neutral Earth supply must be used If the integrity of the external protective conductor...

Страница 18: ...Pumps When the Workstation is initially switched on or when re connected to the AC whilst operating from the internal battery a small delay in the application of AC power will occur between each MDI...

Страница 19: ...on Barcode Reader Barcode LED Primary Speaker Status See Barcode Reader Docking Pump Warning LED N A Status See Docking a Pump and Removing a Pump Nurse Call Nurse Call N A Nurse Call See Nurse Call P...

Страница 20: ...FI signal is a continuous tone Nurse Call Interface A nurse call interface is provided which is activated whenever any Pump attached to the Workstation enters an alarm state and stops when the conditi...

Страница 21: ...smatched Pump alarm location beacon behaviour when using Pumps with selectable alarm profiles Caution If the Pump Alarm Location Beacon fails to illuminate when the Workstation is switched on suspect...

Страница 22: ...601 1 2 and ETSI EN 301 489 17 where applicable If however the Workstation interacts with other equipment measures should be taken to minimise the effects for instance by repositioning or relocation T...

Страница 23: ...wing conditions occur with the Workstation remove it from service for inspection by Qualified Service Personnel dropped excessive moisture fluid spillage high humidity high temperature suspicion of da...

Страница 24: ...ndle is in full working order before use Warning The pole clamp should be used to mount the Workstation on fixed poles overhead swing arm poles only Warning People of a weak disposition should not att...

Страница 25: ...extinguished Verify the System Fault indicator illuminates briefly emits a series of beeps and is then extinguished 9 Nurse call test when connected Nurse call automatically activated Not applicable...

Страница 26: ...once on the auxiliary Workstation to switch the auxiliary Workstation on Removing an Auxiliary Workstation If the primary Workstation cannot be switched off for whatever reason then the Pumps fitted...

Страница 27: ...AC power and data communications once correctly located on the MDI tile InorderfortheWorkstationandPumptocommunicatecorrectly thePumpmustbeconfiguredtoIrDACommunication not RS232 Communication Please...

Страница 28: ...llows the use of either fixed or dynamic network addressing of the Workstation Similarly a DNS client is provided Configuration of these client services is through the Workstation Web Browser User Int...

Страница 29: ...atus of the reader All LEDs off The barcode reader is not receiving power or the barcode reader is in standby mode Steady Green The laser is active Steady Green Single Red Flash The barcode has been r...

Страница 30: ...h BD supplied barcode reader only Workstation Connector Type Binder 09 0998 00 05 Cable Type N A Description 1 2 3 4 5 Pin 1 5V Pin 2 TxD Pin 3 GND Pin 4 RxD Pin 5 SENSE Serial RS232 Interface Connect...

Страница 31: ...Features IEEE 802 11 a b g n Connects to any existing 2 4 or 5 GHz network Wi Fi Protected Access WPA WPA2 and WEP encryption Note If you are upgrading from legacy Workstation to Workstation version...

Страница 32: ...Up to 16 hours to 95 charge time based on battery life cycle and an 11 Pump load Operating time 60 minutes Environmental Operating Transport and Storage Temperature 5 C 40 C 20 C 50 C Humidity 20 90...

Страница 33: ...ibility Configuration1 Trolley Compatible 80083UN00 00 5 Alaris Trolley Compatible 80203UNS00 00 6 80300UNSxx 30 80300UNSxx 32 80300UNSxx 33 80300UNSxx 34 80300UNSxx 50 80300UNSxx 52 80300UNSxx 53 803...

Страница 34: ...or identification of parts Inspect AC outlets communication connectors and the AC inlet for damage Perform electrical safety checks The complete unit leakage current must be measured If more than 500...

Страница 35: ...excess fluid build up on the Workstation Do not use aggressive cleaning agents as these may damage the exterior surface of the Workstation Do not steam autoclave ethylene oxide sterilise or immerse t...

Страница 36: ...to obtain login details Part Number Description 0000EL00889 AC Power Lead UK 0000EL00938 AC Power Lead European Accessories Part Number Description 1000SP00655 Rail Mounting Kit 1000SP00169 Pole Moun...

Страница 37: ...39 0 55 303391 customer service italy bd com Magyarorsz g 36 1 488 0233 info cfn export bd com Nederland 31 0 20 582 94 20 info benelux bd com New Zealand Freephone 0800 572 468 NZ_customerservice bd...

Страница 38: ...nt contains proprietary information of Becton Dickinson and Company or one of its affiliates and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sel...

Отзывы: