Baylis Medical NRG-E-56-32-C0 Скачать руководство пользователя страница 1

 

 

 

Instructions for Use

 

NRG

®

 Transseptal Needle

 

 

ENGLISH ................................................................................................................................................................................................................. 0 
FRANÇAIS ............................................................................................................................................................................................................... 1 
DEUTSCH ................................................................................................................................................................................................................ 3 
NEDERLANDS ......................................................................................................................................................................................................... 5 
ITALIANO ................................................................................................................................................................................................................ 6 
ESPAÑOL ................................................................................................................................................................................................................ 8 
PORTUGUÊS ..........................................................................................................................................................................................................10 
SLOVENSKY ..........................................................................................................................................................................................................11 

 

 

 

 

© Copyright Baylis Medical Company Inc., 2007-2016 

 

The Baylis Medical logo, NRG and TorFlex are 

trademarks and/or registered trademarks of Baylis Medical Company Inc.  

in the United States of America and/or other countries. 

 

Patents pending and/or issued 

 

English 

 

 

Carefully read all instructions prior to use.  Observe all contraindications, warnings and precautions noted in these instructions.  Failure to do so may result in 
patient complications. 
 
CAUTION: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN 
 

I. 

DEVICE

 

DESCRIPTION

 

 

The NRG Transseptal Needle delivers radiofrequency (RF) power in a monopolar mode between its distal electrode and a commercially available external 
Disposable  Indifferent  (Dispersive)  Patch  (DIP) Electrode,  which  is  in  compliance  with IEC  60601-2-2.   The  NRG Transseptal  Needle  is loaded  through  a 
Transseptal  Sheath/Dilator  set,  and  is  connected  at  its  proximal  end  to  the  BMC  Radiofrequency  Puncture  Generator  via  the  BMC  Connector  Cable  and 
optionally to an external pressure monitoring system via a luer connection.  Detailed information concerning the BMC Radiofrequency Puncture Generator is 
contained  in  a  separate  manual  that  accompanies  the  Generator  (entitled  “BMC  Radiofrequency  Puncture  Generator  Instructions  for  Use”).  Generators 
compatible with the NRG Transseptal Needle include the RFP-100 or RFP-100A. 
 
The dimensions for the NRG Transseptal Needle can be found on the device label.  The distal end of the needle contains a hole to facilitate injection of contrast 
solution and the monitoring of cardiac pressures.  As well, the active tip is specially shaped to be atraumatic to the cardiac tissue unless RF energy is applied. 
 

II. 

INDICATIONS

 

FOR

 

USE

 

 

The NRG Transseptal Needle is used to create an atrial septal defect in the heart.  Secondary indications include monitoring intracardiac pressures, sampling 
blood, and infusing solutions. 

III. 

CONTRAINDICATIONS

  

 

The NRG Transseptal Needle is not recommended for use with any conditions that do not require cutting or coagulation of soft tissue. 

 
IV. 

WARNINGS

 

 

  Only physicians with a thorough understanding of angiography and percutaneous interventional procedures should use this device. 

  Do not alter this device in any way. 

  The NRG Transseptal Needle is supplied STERILE using an ethylene oxide process.  Do not use if the package is damaged. 

  Laboratory staff and patients can undergo significant x-ray exposure during radiofrequency puncture procedures due to the continuous usage of fluoroscopic 

imaging.  This exposure can result in acute radiation injury as well as increased risk for somatic and genetic effects. Therefore, adequate measures must be 
taken to minimize this exposure. 

  The NRG Transseptal Needle is intended for single patient use only.  Do not attempt to sterilize and reuse the needle.  Reuse can cause the patient injury 

and/or the communication of infectious disease(s) from one patient to another. Failure to do so may result in patient complications. 

  The NRG Transseptal Needle must be used with the BMC Connector Cable. Attempts to use it with other connector cables can result in electrocution of the 

patient and/or operator. 

  For RFP-100: Do not attempt to puncture with an initial power setting of greater than 10 Watts.  The initial attempt should be made with a setting of 10 Watts. 

In subsequent punctures, the power setting can be increased, if necessary. 

  The pressure transducer system used with the NRG Transseptal Needle must comply with the electrical safety requirements of IEC 60601.  Failure to use 

compliant pressure transducers may result in patient or operator injury. 

 

V. 

PRECAUTIONS

 

 

  Do not attempt to use the NRG Transseptal Needle or ancillary equipment before thoroughly reading the accompanying Instructions for Use. 

  Radiofrequency puncture procedures should be performed only by physicians thoroughly trained in the techniques of radiofrequency powered puncture in a 

fully equipped catheterization laboratory. 

  The sterile packaging should be visually inspected prior to use to detect any compromise.  Ensure that the packaging has not been damaged.  Do not use 

the equipment if the packaging has been compromised. 

  Visually inspect the needle prior to use.  Do not use the needle if there is any damage. 

  Do not use the NRG Transseptal Needle after the “Use By” date indicated on the label. 

  The NRG Transseptal Needle is intended for use with only those devices listed in section VII “Equipment Required.” 

  Read and follow the manufacturer’s instructions for use of the Disposable Indifferent (Dispersive) Patch (DIP) electrode.  Always use DIP electrodes that 

meet or exceed IEC 60601-2-2 requirements. 

  Placement of the dispersive electrode on the thigh or hip could be associated with higher impedance. 

  In order to prevent the risk of ignition make sure that flammable material is not present in the room during RF power application. 

  Take precautions to limit the effects that the electromagnetic interference (EMI) produced by the Generator may have on the performance of other equipment. 

Check the compatibility and safety of combinations of other physiological monitoring and electrical apparatus to be used on the patient in addition to the 
Generator. 

  Adequate filtering must be used to allow continuous monitoring of the surface electrocardiogram (ECG) during radiofrequency power applications. 

  Careful needle manipulation must be performed to avoid cardiac damage, or tamponade.  Needle advancement should be done under fluoroscopic guidance.  

If resistance is encountered, DO NOT use excessive force to advance or withdraw the needle. 

  Do not attempt to puncture until firm position of the active tip has been achieved against the atrial septum. 

  It is not recommended to exceed five (5) radiofrequency power applications per NRG Transseptal Needle. 

  Do not bend the NRG Transseptal Needle.  Excessive bending or kinking of the needle shaft may damage the integrity of the needle and may cause patient 

injury.  Care must be taken when handling the needle. 

  The  Generator  is capable  of  delivering  significant  electrical  power.    Patient  or  operator  injury  can  result from  improper  handling  of the  needle  and  DIP 

electrode, particularly when operating the device. 

  During power delivery, the patient should not be allowed to come in contact with ground metal surfaces. 

  Apparent low power output or failure of the equipment to function properly at normal settings may indicate faulty application of the DIP electrode, failure to 

an electrical lead, or poor tissue contact at the active tip.  Check for obvious equipment defects or misapplication.  Attempt to better position the tip of the 
needle against the atrial septum.  Only increase the power if low power output persists. 

  Baylis Medical Company relies on the physician to determine, assess and communicate to each individual patient all foreseeable risks of the Baylis Medical 

Radiofrequency Puncture System. 

 

VI. 

ADVERSE

 

EVENTS

 

 

Adverse events that may occur while using the Baylis Medical Radiofrequency Puncture System include: 

Tamponade 

Sepsis/Infection 

Thromboembolic episodes 

Vessel perforation 

Atrial Fibrillation 

Myocardial Infarction 

Vessel spasm 

Sustained arrhythmias 

Atrial Flutter 

Hemorrhage 

Vascular thrombosis 

Perforation of the myocardium 

Hematoma 

Allergic reaction to contrast medium 

Ventricular Tachycardia 

Pain and Tenderness 

Arteriovenous fistula 

 

 

VII. 

EQUIPMENT

 

REQUIRED 

Intracardiac  puncture  procedures  should  be  performed  in  a  specialized  clinical  setting  equipped  with  a  fluoroscopy  unit,  radiographic  table,  physiologic 
recorder, emergency equipment and instrumentation for gaining vascular access.  Ancillary materials required to perform cardiac Puncture include: 

 

  BMC Radiofrequency Puncture Generator 

  Baylis Connector Cable (RFP-102 or RFP-103 (model dependent for NRG Transseptal needle) for use with RFP-100 Generator, or RFX-BAY-TS for use 

with RFP-100A Generator). 

  Transseptal Sheath/Dilator kit, such as the Baylis Medical Company 

TorFlex

TM

 

Transseptal Guiding Sheath. 

  Disposable Indifferent (dispersive) Patch (DIP) electrode must meet or exceed IEC 60601-2-2 requirements for electrosurgical electrodes. 

 

 

Baylis Medical Company Inc. 
5959 Trans-Canada Highway 
Montreal, Quebec, Canada, H4T 1A1 
Tel: (514) 488-9801/ (800) 850-9801 Fax: (514) 488-7209 
www.baylismedical.com 

 

 

 

 

 

EU Authorized Representative: 

Quality First International 
20 Eversley Road, Bexhill-on-Sea, East Sussex 
TN40 1HE, United Kingdom 
Tel: +44-(20)-8-522-1937 Fax: +44-(20)-8-522-1937 

1 of 7 

 

DMR NRG 3.3 (IFU) V-11 23-Nov-2018 

VIII. 

INSPECTION

 

PRIOR

 

TO

 

USE 

Prior to use of the Baylis Medical Radiofrequency Puncture System, the individual components including the BMC Radiofrequency Puncture Generator, NRG 
Transseptal Needle, and the BMC Connector Cable should be carefully examined for damage or defects, as should all equipment used in the procedure.  Do 
not use defective equipment. 
 

IX. 

DIRECTIONS

 

FOR

 

USE

 

 

  All instructions for equipment required should be carefully read, understood, and followed.  Failure to do so may result in complications. 

  The NRG Transseptal Needle is supplied sterile.  Use aseptic technique when opening the package and handling the product in the sterile field. 

  Thoroughly flush the NRG Transseptal Needle with heparinized saline solution prior to use. 

  A Transseptal Sheath and Dilator are usually inserted through the right femoral vein and are then advanced over a guidewire to be positioned into the superior 

vena cava (SVC) under fluoroscopic guidance.  The Baylis Medical TorFlex Transseptal Guiding Sheath is recommended for this purpose. 

  Insert the NRG Transseptal Needle through the sheath/dilator set until the tip of the needle is just within the dilator.  Ensure the needle is free to twist and/or 

rotate without resistance, as it is advanced to this position. 

  If using a pressure monitoring system, connect the NRG Transseptal Needle to it by joining its luer connector on the handle to a luer lock and rotating the 

connector to ensure a secure connection. 

  Connect the NRG Transseptal Needle to the BMC Connector Cable. Make sure that the Connector Cable is plugged into the appropriate port on the BMC 

Radiofrequency Puncture Generator.  Be sure to carefully follow the Instructions for Use provided with the Generator and Cable. 

  Position the tip of the needle/sheath/dilator assembly so that the dilator is engaging the septum at the fossa ovalis under fluoroscopic guidance. 

  Deliver radiofrequency power via the BMC Radiofrequency Puncture Generator and advance the NRG Transseptal Needle through the septum into the left 

atrium.  Please refer to the Generator Instructions for Use before using the Generator. 

 

NOTE: It is recommended that the user use the least amount of energy to achieve the desired puncture.

 

  For RFP-100: A power setting of 10 Watts has been experimentally determined to be sufficient for successful puncture. 

  For RFP-100A: An initial RF setting between one (1) second on “PULSE” mode to two (2) seconds on “CONSTANT” mode has been shown to be sufficient 

for successful puncture. 

  Radiofrequency power delivery can be terminated by pressing the RF ON/OFF button on the Generator if the timer has not expired. 

  Entry into the left atrium can be confirmed using fluoroscopy.  Further confirmation can be obtained by either observing a left atrial pressure tracing, by 

injecting a small amount of contrast media through the needle, or by aspiration of blood. 

  If septal puncture is not successful after five (5) radiofrequency power applications, it is advised that the user proceed with an alternate method for the 

procedure. 

  Once successful puncture into the left atrium is confirmed, the NRG Transseptal Needle may be carefully advanced without any radiofrequency power. 

  The transseptal dilator can be advanced over the needle to enlarge the puncture. 

  Remove the NRG Transseptal Needle slowly. 

 

Connections

 (pg.07) 

 

X. 

CLEANING

 

AND

 

STERILIZATION

 

INSTRUCTIONS 

The NRG Transseptal Needle is intended for single use only.  Do not clean or re-sterilize the NRG Transseptal Needle. 
 

XI. 

TROUBLESHOOTING

 

 

The following table is provided to assist the user in diagnosing potential problems. 
 

PROBLEM 

COMMENTS 

TROUBLESHOOTING 

Generator 
Error 
Messages 

In order to successfully 
puncture tissue using 
RF energy, the entire 
system must be 
connected and all 
devices must be in good 
working order. 

Ensure that all connections are made: 
- needle to connector cable 
- connector cable to generator 
- generator to power outlet 
- generator to grounding pad 
 
Visually inspect the needle or cable for damage.  Immediately discard any damaged equipment.  If 
the problem persists discontinue use. 
 
For error messages encountered while attempting radiofrequency puncture, refer to the operator’s 
manual that accompanies the Generator. 

Inaccurate 
Pressure 
Readings 

In order to accurately 
monitor pressure, the 
entire system must be 
properly connected and 
all devices must be in 
good working order. 

Ensure that the following connections are made: 
-needle to pressure transducer 
-pressure transducer to monitoring system 
 

  Ensure that the transducer is zeroed. 

  Ensure that the transducer is leveled with the phlebostatic axis 

  Perform a ‘fast-flush test’ to determine the dynamic response. 

  Visually inspect the needle for any damage.  If there are any breaks or kinks, discard immediately 

Needle 
breaks or 
kinks. 

Breaks and kinks in the 
needle are a potential 
cause of patient injury. 

Discard immediately 

 

XII. 

CUSTOMER

 

SERVICE

 

AND

 

PRODUCT

 

RETURN

 

INFORMATION

 

 

If you have any problems with or questions about Baylis Medical Equipment contact our technical support personnel.  

NOTES: 

1. In order to return products you must have a return authorization number before shipping the products back to Baylis Medical Company. 
2. Baylis Medical will not accept any piece of used equipment without a sterilization certificate.  Ensure that any product being returned to Baylis Medical has 

been cleaned, decontaminated and sterilized as indicated in the user instructions before returning it for warrantied service. 

XIII. 

LABELING AND SYMBOLS 

 

Manufacturer 

 

Caution:

 Federal (U.S.A.) law restricts this device to sale by or on the order 

of a physician. 

EU Authorized Representative 

 

Single Use – Do not reuse 

Sterile using ethylene oxide 

 

Lot Number 

 

Use By 

 

Only for EU member states:

Use of this symbol indicates that the product must be disposed of in a way 
that complies with local and national regulations. For questions regarding 
recycling of this device please contact your distributor

 

 

Caution 

 

Do Not Use if Packaging is Damaged

 

 

Consult Instructions for Use  

 

Keep Away from Sunlight 

 

Model number  

 

Non-pyrogenic 

 

XIV. 

LIMITED

 

WARRANTY

 

 

D

ISPOSABLES AND 

A

CCESSORIES

 

Baylis Medical Company Inc. (BMC) warrants its Disposable and Accessory products against defects in materials and workmanship.  BMC warrants that sterile 
products will remain sterile for a period of time as shown on the label as long as the original package remains intact. Under this Limited Warranty, if any covered 
product is proved to be defective in materials or workmanship, BMC will replace or repair, in its absolute and sole discretion, any such product, less any charges 
to BMC for transportation and labor costs incidental to inspection, removal or restocking of product. The length of the warranty is: (i) for the Disposable products, 
the shelf life of the product, and (ii) for the Accessory products, 90 days from shipment date. This limited warranty applies only to new original factory delivered 
products that have been used for their normal and intended uses.  BMC’s Limited Warranty shall not apply to BMC products which have been resterilized, 
repaired, altered, or modified in any way and shall not apply to BMC products which have been improperly stored or improperly cleaned, installed, operated or 
maintained contrary to BMC’s instructions. 
 

DISCLAIMER AND LIMITATION OF LIABILITY  
 

THE  LIMITED  WARRANTY  ABOVE  IS  THE  SOLE  WARRANTY  PROVIDED  BY  SELLER.  SELLER  DISCLAIMS  ALL  OTHER  WARRANTIES, 

WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE OR PURPOSE.   
 

THE  REMEDY  SET  FORTH  HEREIN  SHALL  BE  THE  EXCLUSIVE  REMEDY  FOR  ANY  WARRANTY  CLAIM,  AND  ADDITIONAL  DAMAGES, 

INCLUDING  CONSEQUENTIAL  DAMAGES  OR  DAMAGES  FOR  BUSINESS  INTERRUPTION  OR  LOSS  OF  PROFIT,  REVENUE,  MATERIALS, 
ANTICIPATED SAVINGS, DATA, CONTRACT, GOODWILL OR THE LIKE (WHETHER DIRECT OR INDIRECT IN NATURE) OR FOR ANY OTHER FORM 
OF INCIDENTAL, OR INDIRECT DAMAGES OF ANY KIND, SHALL NOT BE AVAILABLE. SELLER'S MAXIMUM CUMULATIVE LIABILITY RELATIVE TO 
ALL OTHER CLAIMS AND LIABILITIES, INCLUDING OBLIGATIONS UNDER ANY INDEMNITY, WHETHER OR NOT INSURED, WILL NOT EXCEED THE 
COST  OF  THE  PRODUCT(S)  GIVING  RISE  TO  THE  CLAIM  OR  LIABILITY.  SELLER  DISCLAIMS  ALL  LIABILITY  RELATIVE  TO  GRATUITOUS 
INFORMATION  OR  ASSISTANCE  PROVIDED  BY,  BUT  NOT  REQUIRED  OF  SELLER  HEREUNDER.  ANY  ACTION  AGAINST  SELLER  MUST  BE 
BROUGHT WITHIN EIGHTEEN (18) MONTHS AFTER THE CAUSE OF ACTION ACCRUES. THESE DISCLAIMERS AND LIMITATIONS OF LIABILITY 
WILL  APPLY  REGARDLESS  OF  ANY  OTHER  CONTRARY  PROVISION  HEREOF  AND  REGARDLESS  OF  THE  FORM  OF  ACTION,  WHETHER  IN 
CONTRACT,  TORT  (INCLUDING  NEGLIGENCE  AND  STRICT  LIABILITY)  OR  OTHERWISE,  AND  FURTHER  WILL  EXTEND  TO  THE  BENEFIT  OF 
SELLER'S VENDORS, APPOINTED DISTRIBUTORS AND OTHER AUTHORIZED RESELLERS AS THIRD-PARTY BENEFICIARIES. EACH PROVISION 
HEREOF WHICH PROVIDES  FOR  A  LIMITATION  OF  LIABILITY,  DISCLAIMER  OF WARRANTY  OR  CONDITION  OR EXCLUSION  OF  DAMAGES  IS 
SEVERABLE AND INDEPENDENT OF ANY OTHER PROVISION AND IS TO BE ENFORCED AS SUCH.

 

 

IN ANY CLAIM OR LAWSUIT FOR DAMAGES ARISING FROM ALLEGED BREACH OF WARRANTY, BREACH OF CONTRACT, NEGLIGENCE, 

PRODUCT LIABILITY OR ANY OTHER LEGAL OR EQUITABLE THEORY, THE BUYER SPECIFICALLY AGREES THAT BMC SHALL NOT BE LIABLE 
FOR DAMAGES OR FOR LOSS OF PROFITS, WHETHER FROM BUYER OR BUYER’S CUSTOMERS.  BMC’S LIABILITY SHALL BE LIMITED TO THE 
PURCHASE COST TO BUYER OF THE SPECIFIED GOODS SOLD BY BMC TO BUYER WHICH GIVE RISE TO THE CLAIM FOR LIABILITY. 

 

No agent, employee or representative of Baylis Medical has the authority to bind the Company to any other warranty, affirmation or representation 

concerning the product. 
 

This warranty is valid only to the original purchaser of Baylis Medical products directly from a Baylis Medical authorized agent. The original purchaser 

cannot transfer the warranty. 
 

Use of any BMC product shall be deemed acceptance of the terms and conditions herein. 

 

The warranty periods for Baylis Medical products are as follows: 

Disposable Products 

The shelf life of the product 

Accessory Products 

90 days from the shipment date 

 

Français

 

 

Lire attentivement toutes les directives avant l’utilisation. Respecter toutes les contre-indications, avertissements et précautions indiqués dans ces directives. 
Leur non-respect risque de causer des complications pour le patient. 
AVERTISSEMENT : EN VERTU DE LA LOI FÉDÉRALE AMÉRICAINE, CE DISPOSITIF NE PEUT ÊTRE VENDU QUE PAR UN MÉDECIN OU SUR L’AVIS 
D’UN MÉDECIN 
 

I. 

DESCRIPTION

 

DU

 

DISPOSITIF 

NRG Transseptal Needle libère une énergie radiofréquence (RF) en mode monopolaire entre son électrode distale et une électrode indifférente de retour à 
utilisation unique offerte sur le marché, laquelle doit être conforme aux normes IEC 60601-2-2. NRG Transseptal Needle s’insère dans une gaine transseptale/un 
dispositif de dilatation; son extrémité proximale se branche au BMC Radiofrequency Puncture Generator  par le BMC Connector Cable ; elle peut également 
être connectée à un appareil de mesure de la pression externe au moyen d’un raccord Luer. Vous trouverez de l’information détaillée sur le BMC Radiofrequency 
Puncture  Generator    dans  le  manuel  distinct  accompagnant  le  générateur  (intitulé  « Mode  d’emploi  du  BMC  Radiofrequency  Puncture  Generator   »). 
Générateurs compatibles avec  NRG Transseptal needle comprennent la RFP-100 ou RFP-100A. 
 
Les dimensions de NRG Transseptal Needle se trouvent sur l’étiquette du dispositif. L’extrémité distale de l’aiguille est dotée d’un orifice afin de faciliter l’injection 
d’une solution de contraste et la surveillance des pressions cardiaques. Par ailleurs, l’extrémité active est spécialement conçue de façon à ne pas causer de 
lésions aux tissus cardiaques lorsqu’il n’y a pas administration d’énergie RF. 

II. 

INDICATIONS

 

D’EMPLOI

 

 

NRG Transseptal Needle est utilisée pour créer une communication interauriculaire dans le cœur. Les indications secondaires comprennent la surveillance des 
pressions intracardiaques, le prélèvement sanguin et la perfusion de solutions. 

Содержание NRG-E-56-32-C0

Страница 1: ...ector Cable is plugged into the appropriate port on the BMC Radiofrequency Puncture Generator Be sure to carefully follow the Instructions for Use provided with the Generator and Cable Position the ti...

Страница 2: ...estions concernant des produits de Baylis M dical communiquez avec notre quipe de support technique REMARQUES 1 Pour retourner un produit il vous faut un num ro d autorisation de retour avant de l exp...

Страница 3: ...N DAR BER HINAUS F R DIE H NDLER DIE ERNANNTEN DISTRIBUTOREN UND SONSTIGE AUTORISIERTE WIEDERVERK UFER DES VERK UFERS ALS DRITTE BEG NSTIGTE JEDE BESTIMMUNG HIERIN DIE EINE BESCHR NKUNG DER HAFTUNG EI...

Страница 4: ...sterilizzare e a riutilizzare l ago Riusarlo potrebbe causare ferite al paziente e o trasmettere malattie infettive da un paziente all altro La loro inosservanza pu provocare complicanze nei pazienti...

Страница 5: ...en darse mientras se utiliza el Baylis Medical Radiofrequency Puncture System se incluyen Taponamiento Septicemia Infecci n Episodios tromboemb licos Perforaci n vascular Fibrilaci n atrial Infarto de...

Страница 6: ...RF O dilatador transeptal pode ser guiado sobre a agulha para alargar a pun o Retire lentamente a NRG Transseptal Needle Conex es p gina 07 X INSTRU ES DE LIMPEZA E ESTERILIZA O A NRG Transseptal Need...

Страница 7: ...er tor k elektrickej z suvke gener tor k uzem ovacej elektr de Vizu lne skontrolujte i nie je po koden ihla alebo k bel Ak ko vek po koden zariadenie okam ite zlikvidujte Ak probl m pretrv va zariaden...

Отзывы: