Bayer Customer Ser
vice: (852) 8
100 6386
36
37
www
.bayerdiabetes.com
W
ARNING: KEEP BA
T
TER
Y A
W
A
Y
FROM
CHILDREN! LITHIUM BA
TTERIES ARE POISONOUS!
If swallowed, immediately contact your physician or
poison contr
ol center
.
Dispose of old batteries pr
operly
.
C
LEANING
Y
OUR
M
ETER
:
Your
щଥ
(
Bai An Jie
) meter can be cleaned using
a moist
(not wet)
lint-fr
ee tissue with a mild deter
gent or
disinfecting solution (1 part bleach mixed with 9 parts water).
Do not use alcohol.
Wipe the
outside of the meter
, taking car
e
to keep deter
gent or disinfecting
solution fr
om running down into
the test strip slot. Wipe dry with
a lint-fr
ee tissue after cleaning.
If you do get moistur
e in the
test strip slot, wick it away with
a corner of a tissue as shown.
Inspect the test strip slot, fr
om
both the inside and outside, for
debris and foil. Gently r
emove
if found.
NOTE:
When the battery life is too low to perform a test,
the display will show only a battery symbol. When this
occurs, the meter will not allow you to run a test until the
battery is changed. When the battery life has completely
ended, nothing will show on the display and the meter
will not r
espond.
CAUTION:
If you have a low
battery in your meter
, you
should r
eplace the battery
within appr
oximately 20
readings or one week.
R
EPLACING
THE
B
A
TTER
Y
:
Your
щଥ
(Bai An Jie
) meter uses one 3-volt lithium
battery (CR2032). A battery will alr
eady be installed in the
meter when you buy it.
A fl
ashing battery on the scr
een
means that it is time to put a
new battery in the meter
. Y
ou
will not erase test r
esults in
the meter’
s memory when you
change the battery
. Y
ou may
need to r
eset the time.
W
ARNING: Potential Biohazard!
Healthcar
e pr
ofessionals should be awar
e of the
following
and should follow the infection contr
ol
pr
ocedur
e appr
oved by their facility
.
All pr
oducts or objects that come into contact with
human blood, even after cleaning, should be handled as
if capable of transmitting viral diseases.
The user should follow the r
ecommendations for pr
evention of blood-borne
transmissible diseases in healthcar
e settings, as r
ecommended for potentially
infectious human blood specimens in National Committee for Clinical Labora-
tory Standar
ds, Pr
otection of Laboratory W
orkers fr
om Occupationally Acquir
ed
Infections: Appr
oved Guideline
–
Second Edition. NCCLS Document M29-A2 [ISBN
1-56238-453-8] NCCLS, 940 W
est V
alley Road, Suite 1400, W
ayne, Pennsylvania
19087-1898, USA, 2001. This document has complete information on the topic of
user pr
otection and can be used as backgr
ound material for instruction.
82076779_BRZ2_UG_EN_FpBp_v0.indd 36-3782076779_BRZ2_UG_EN_FpBp_v0.indd 36-37
8/8/11 12:45:45 PM8/8/11 12:45:45 PM