16
For example:
When the Min. display value selected is 1 and the Max. capacity selected is 60kg, the graduation number
to be accepted is 6000, neither 600 nor 60000. And when the Min. display value and the Max. capacity
you selected is separately 1 and 1000 kg, the graduation number to be accepted is 10000, neither 1000
nor 100000.
16.3.2 After you press (ENTER) to confirm the Max. capacity, the indicator displays and
carry through countdown, at this time it automatically confirm zero position.
16.3.3 After finishing the countdown, it will display . After confirmation by pressing (ENTER)
key, the display of the indicator has 2 possibilities.
a. When F1.2=1, wich means the case your selection is fixed point calibration, at this time the
indicator displays a weight data. See the following table for the relation between the weight data
for rectifying and the Max. Capacity.
ENTER
T
ENTER
T
6
/12
15
/20
30
/60
60
/120
150
/300
300
/600
600
/1200
1000
/2000
1500
/3000
2000
/4000
5000
/10000
3000
/6000
10000
/20000
Max.
Capacity
(kg/lb)
Rectifying
Value
(kg/lb)
5
/10
10
/20
20/50
100/200
200/500
600/1200
1000/2000
3000
/5000
At this moment you can load test weights (or substitute) on the platform and press (ENTER), the
indicator will display and then carry through countdown. After the countdown is finished, the
whole calibration is finished, at this time the indicator displays . You can exit by pressing
(EXIT).
b. When F1.2=2 or F1.2=3 (which means the case your selection is arbitrary point calibration),
the indicator displays . Under this state, you can load the test weights (or substitute) on
the platform (generally the weight loaded should be more than 60% Max. capacity to increase
linearity), and press to make it display , the decimal point will appear in the
corresponding position and the last “0” will flash. Input the data equal to loaded weight through
cooperative use of and as interpreted in item 5 and press (ENTER) to confirm.
The indicator will display and countdown. When the countdown is finished, it indicates
the calibration is over and the indicator displays . Press (EXIT) to exit from
calibrating state and make the indicator self-check. After self-check, the indicator will back to
normal weighing state.
ENTER
T
C
EXIT
BL
0
UNITS
0
ENTER
T
C
EXIT
BL
When the version number is less than REV2.0, you can only choose arbitrary
point calibration.
16.5
Logic flow chart for setting and calibrating
USER’S MANUAL
CWP. Bench scale