F2.3 Digital filtering coefficient selecting, with 3 options
F2.3=1 low-degree filtering: suitable for the conditions with weak air flow or little atmospheric
pressure charge.
F2.3=2 medium-degree filtering: suitable for general occasion.
F2.3=3 high-degree filtering: suitable for the situations with strong air flow or large atmospheric
pressure change.
F2.4 Extension display mode (i.e. Internal graduation number display), with 2 options
F2.4=0 which means prohibiting extension display
F2.4=1 which means allowing extension display
Note:
“extension display” refers to internal resolution ration display, you can use this method
conveniently to adjust and view the error.
F2.5: selecting for setup value from F2.1 to F2.4, with 2 options
F2.5=1 selection of setup value of the users.
F2.5=2 selection of default value, refer to the following table for the data
F2.1
20
F2.2
0.5
F2.3
2
F2.4
0
16.2.3 Subdirectories and contend about F3
F3 is used for calibrating
16.3 Calibration
16.3.1 When the indicator displays , press (ENTER) key to enter calibrating state. After
entering, the indicator will firstly display , which indicates Min. Display value. There are 3
options for the Min. Display value, and the value is separately 1, 2 and 5. You can select Min display
value via , and then press (ENTER) to confirm.
ENTER
T
UNITS
ENTER
T
If the version number is less than REV2.0, then the Max Capacity must be
through digit keys.
The steps are as following:
1.
Digit input
a. When the Min. Display value is confirmed, the indicator will display ,
is the last Max. Capacity. If you want to change it, just press key to make it
zero, that’s . And the decimal point disappers.
b. Input the values you need with methods interpreted in item 5.
2.
Decimal point input.
a. Press (I/O) and (ENTER) to make the decimal point stay at the right but one, that’s
is the last input value.
b. Press key to make the decimal point stay in the position as you need.
4
3
2
1
0
4
3
2
1
0
0
ENTER
T
I/O
POWER
4
3
2
1
0
4
3
2
1
0
0
If the version number is less than REV2.0, the Max Capacity can be pre-setted.
After confirmed, the indicator will display . The value display represents Max. Weighing
capacity, and there are 13 options for Max. capacity, reparately: 6kg (12lb), 15kg (30lb), 30kg (60lb), 60kg
(120lb) 150kg (300lb), 300kg (600lb), 600kg (1200lb), 1000kg (2000lb), 1500kg (3000lb), 2000kg
(4000lb), 3000kg (6000lb), 5000kg (1000lb), 10000kg (20000lb).
You can select Max. xeighing capacity as you need by pressing .
Notice:
The scale (indicator) is able to comprehend graduation number automatically through the Min.
display value and Max. capacity you select. And the graduation number range to be comprehened is
always within 2000~15000.