F2.3 Digital filtering coefficient selecting, with 3 options

F2.3=1 low-degree filtering: suitable for the conditions with weak air flow or little atmospheric

pressure charge.

F2.3=2 medium-degree filtering: suitable for general occasion.
F2.3=3 high-degree filtering: suitable for the situations with strong air flow or large atmospheric

pressure change.

F2.4 Extension display mode (i.e. Internal graduation number display), with 2 options

F2.4=0 which means prohibiting extension display
F2.4=1 which means allowing extension display

Note:

 “extension display” refers to internal resolution ration display, you can use this method 

conveniently to adjust and view the error.

F2.5: selecting for setup value from F2.1 to F2.4, with 2 options

F2.5=1 selection of setup value of the users.
F2.5=2 selection of default value, refer to the following table for the data

F2.1

20

F2.2

0.5

F2.3

2

F2.4

0

16.2.3 Subdirectories and contend about F3

F3 is used for calibrating

16.3 Calibration

16.3.1 When the indicator displays                , press      (ENTER) key to enter calibrating state. After 
entering, the indicator will firstly display                 , which indicates Min. Display value. There are 3 
options for the Min. Display value, and the value is separately 1, 2 and 5. You can select Min display 
value via      , and then press      (ENTER) to confirm.

ENTER

T

UNITS

ENTER

T

If the version number is less than REV2.0, then the Max Capacity must be
through digit keys.

The steps are as following:

1.

Digit input
a. When the Min. Display value is confirmed, the indicator will display                       ,
                      is the last Max. Capacity. If you want to change it, just press          key to make it
    zero, that’s                  . And the decimal point disappers.
b. Input the values you need with methods interpreted in item 5.

2.

Decimal point input.
a. Press      (I/O) and        (ENTER) to make the decimal point stay at the right but one, that’s 

      is the last input value.

b. Press         key to make the decimal point stay in the position as you need.

4

3

2

1

0

4

3

2

1

0

0

ENTER

T

I/O

POWER

4

3

2

1

0

4

3

2

1

0

0

If the version number is less than REV2.0, the Max Capacity can be pre-setted.

After confirmed, the indicator will display                 . The value display represents Max. Weighing 
capacity, and there are 13 options for Max. capacity, reparately: 6kg (12lb), 15kg (30lb), 30kg (60lb), 60kg 
(120lb) 150kg (300lb), 300kg (600lb), 600kg (1200lb), 1000kg (2000lb), 1500kg (3000lb), 2000kg 
(4000lb), 3000kg (6000lb), 5000kg (1000lb), 10000kg (20000lb).
You can select Max. xeighing capacity as you need by pressing       .

Notice:

 The scale (indicator) is able to comprehend graduation number automatically through the Min. 

display value and Max. capacity you select. And the graduation number range to be comprehened is 
always within 2000~15000.

UNITS

USER’S MANUAL

CWP. Bench scale

Содержание CWP

Страница 1: ...CWP Bench scale USER S MANUAL...

Страница 2: ...2 Backlight selecting 13 Auto rectifying 14 Storage battery maintaining 15 Use guide 16 Data setting and calibrating 3 3 4 6 7 7 9 9 9 9 11 12 12 12 13 13 CONTENTS USER S MANUAL CWP Bench scale Discla...

Страница 3: ...gnal input range 0mV 8mV 1 2 Zero point temperature drift 0 15 V C 1 3 Sensitivity temperature drift 10 ppm C 1 4 A D conversion resolution ration 500000 1 5 Display resolution ration 30000 1 6 A D co...

Страница 4: ...ustable support Load cell input 7 core RS 232 Interface 5 core Power supply adaptor Battery switch Seal ring assembly PCB Indicator enclosure Indicator rear cover Storage battery Battery box cover Fig...

Страница 5: ...gle cursor appears here See item 12 3 for details 7 CHARGE indicating symbol it is on when the battery is recharging 8 Auto Off indicating symbol When you switch on the auto shut off function the tria...

Страница 6: ...pressing this key and the cursor on the screen will change from G to N At this time if tou remove the goods from the scale platform the screen displays a negative value Press this key again the displa...

Страница 7: ...At this time if you press the selected falshing digit will change in 0 1 2 9 0 For example to input a group of digit 2345 Step 1 Press C the indicator display Note While inputing the digit first press...

Страница 8: ...oad mark bit D2 0 normality D2 1 overload D1 Positive and negative mark bit D1 0 positive number D1 1 negative number D0 Gross weight and net weight mark bit D0 0 gross weight D0 1 net weight State C...

Страница 9: ...ic weighing mode after power on 7 POWER ON OPERATION I O POWER You can convert between basic weighing and check weighing by pressing MODE key 8 TWO WEIGHING MODES SWITCHING MODE SET After power on the...

Страница 10: ...and low limit value in percentage mode 10 4 1 After the target weight setup is finished and the indicator displays and if the item 10 4 2 can be met i e the absolute value of the value difference betw...

Страница 11: ...splay between value difference display and normal display by pressing key and press once the display will be converted once It can reserve the shift mode unless the power supply is cut off 11 4 Three...

Страница 12: ...will display At this time place a test weight or substitute on scale platform Because the division value and Max capacity are known the decimal point place need not to be considered So at this time o...

Страница 13: ...horisation contact our sales agent in your area 15 7 If you need to re calibrate the indicator do it after the indicator is warm up for 30 minutes 15 USE GUIDE Forbid changing the data of the scale un...

Страница 14: ...s less than REV2 0 you can only choose F1 2 3 F1 5 Serial output mode with 2 options F1 5 0 printing output printer model TM U295 F1 5 1 serial output F1 6 Selecting setup value from F1 1 to F1 5 with...

Страница 15: ...ct Min display value via and then press ENTER to confirm ENTER T UNITS ENTER T If the version number is less than REV2 0 then the Max Capacity must be through digit keys The steps are as following 1 D...

Страница 16: ...t this moment you can load test weights or substitute on the platform and press ENTER the indicator will display and then carry through countdown After the countdown is finished the whole calibration...

Страница 17: ...main directory and subdirectory by Figure 12b Press to enter subdirectory and enter next main directory by confirming the last subdirectory T F1 F1 1 1 F1 2 1 F1 6 1 F2 F2 1 0 F2 F1 1 2 F2 1 0 F1 2 2...

Страница 18: ...ER T UNITS F3 F1 Inputting or selecting Max capacity Selecting decimal point Confirming zero positioning Arbitrary point loading Fixed point loading d 1 Select 1 2 or 5 by Count down SCL 09 F3 Self te...

Страница 19: ...19 USER S MANUAL CWP Bench scale...

Страница 20: ......

Отзывы: