14
16.2.1. Subdirectorios y contenido en F1
F1.1: Unit selecting, with 4 options
F1.1=1 only for choosing “kg” unit.
F1.1=2 “kg” and “lb” unit choosing and converting, when turn-on, its unit is “kg” and can be
calibrated only under “kg” unit.
F1.1=3 “kg” and “lb” unit choosing and converting, when turn-on, its unit is “lb” and can be
calibrated only under “lb” unit.
F1.1=4 only for choosing “lb” unit.
F1.2: Calibration mode selecting, with 3 options
F1.2=1 fixed point calibrating and selecting internal preset Max. capacity.
F1.2=2 arbitrary point calibrating and selecting internal preset Max. capacity.
F1.2=3 arbitrary point calibrating and selecting Max. capacity.
F1.3. Baud rate selecting, with total 5 options
F1.3=1 B=600
F1.3=4 B=4800
F1.3=2 B=1200
F1.3=5 B=9600
F1.3=3 B=2400
F.1.4 Auto shut-off function, with 3 options.
F1.4=0 auto shut-off function off
F1.4=3 auto shut-off in 3 minutes after operation stopped
F1.4=5 auto shut-off in 5 minutes after operation stopped
When the version is less than REV2.0, you can only choose F1.2=3
F1.5 Serial output mode, with 2 options
F1.5=0 printing output (printer model: TM-U295)
F1.5=1 serial output.
F1.6.Selecting setup value from F1.1 to F1.5, with 2 options.
F1.6 selection of setup value of the users
F1.6 selection of default value, refer to the following table
F1.1
1
F1.2
3
F1.3
2
F1.4
1
F1.5
1
*Notice: After the option of F1.6 is finished and if you press (ENTER) to confirm, at this time the
indicator will display again to enter next main directory.
When pressing , you can re-select main directories. When pressing (EXIT), you will exit from the
main directory.
16.2.2 Subdirectories and contend about F2.
F2.1 Turn-on zero range selecting, with 4 options
F2.1=0 auto turn-on zero function off.
F2.1=5 turn-on zero range of 5 % F·S
F2.1=10 turn-on zero range of 10 % F·S
F2.1=20 turn-on zero range of 20 % F·S
F2.2 Auto zero tracking range selecting, with 5 options
F2.2=0.0 no zero tracking.
F2.2=0.2 auto zero tracking of 0.2e
F2.2=0.5 auto zero tracking of 0.5e
F2.2=1.0 auto zero tracking of 1.0e
F2.2=2.0 auto zero tracking of 2.0e
F2.2=3.0 auto zero tracking of 3.0e
ENTER