Message
Flow
Status
Condition
Alarm Messages
CHECK SET
LOADING
No flow
I.V. set is not loaded properly. Open the pump door and ver-
ify that the set is loaded properly against the pumping fin-
gers and through the upstream and downstream occlusion
sensors. Then restart the pump.
AIR
No flow
a. Air bubble at detector
b. Empty fluid container
c. No I.V. set in pump
OCCLUSION
No flow
Closed distal clamp, stopcock, clogged filter or other block-
age downstream of the pump.
Note:
The pump can be set up to restart itself automatically
when the occlusion is removed. If you want to prevent the
pump from restarting automatically, press any enabled key
(such as
SILENCE
) during the alarm.
INSERT SLIDE
CLAMP
Note: This
alarm option is
available only
on pumps
running
software
versions later
than 1.08.
No flow
The slide clamp loading option is enabled and the slide
clamp is not loaded in the slide clamp slot. To clear this
alarm, open the pump door and insert the slide clamp into
the slot. Close the door.
UPSTREAM
OCCLUSION
No flow
Closed clamp or other blockage upstream of the pump.
DOOR OPEN
No flow
Pump door not fully closed. Ensure that the I.V. set is prop-
erly loaded and close the pump door.
BATTERY LOW
with rapid
three-tone
alarm
No flow
Battery power has been exhausted. Plug pump into AC out-
let to restore operation and recharge battery.