d.
To begin the Programmed Delivery
Profile, press
PRI START
. The mes-
sage
REVIEW PDP
appears if you
have not completed reviewing the
stored PDP. After you have reviewed
all steps and pressed
PRI START
, the
pump begins delivering fluid.
Note:
If you enter Programmed Delivery Profile mode and
then try to start the pump without first entering a
PDP, the message
ENTER PGM
appears on the second
line of the message display. The message will disap-
pear once you start to enter a PDP.
4.
Modify previous Pro-
grammed Delivery steps.
a.
Press the
NEXT
key to reach the Pro-
gram Delivery step you wish to
modify.
b.
Press the key for the infusion pa-
rameter you wish to modify (
PRI VTBI
or
PRI RATE
).
c.
Enter the new value on the numeri-
cal keyboard.
d.
Press the
NEXT
key to lock in the pro-
grammed values and advance to the
next step.
e.
Modify additional Program Delivery
steps if desired. After reviewing or
changing the last step, press
NEXT
again, until the first step is dis-
played.
f.
To begin the Program Delivery Pro-
file, press
PRI START
. The pump will
begin delivering fluid.
5.
To program new manual
Program Delivery steps:
23
Rev.B