When the Ramp PDP is running the pump will display the
rate and VTBI of each step, along with the time remaining
until completion of the Ramp PDP. A message will indicate
the current phase of the Ramp PDP (
DELAY, RAMP UP,
LEVEL DELV, RAMP DOWN
).
PDP Phase
Description
Delay Time
This is the time delay from the start of the infusion until the actual
therapy begins. During this phase the pump delivers fluid at 5mL/hr
KVO rate.
Run Time
This represents the total time of the Ramp PDP, excluding the delay
time.
Ramp Up
Time
This represents the time from the start of the therapy until the level
delivery rate is achieved.
Ramp Down
Time
This represents the time to taper off from the level delivery rate. After
completion of the Ramp PDP the pump continues at the KVO rate.
25
Rev.B
rate
time
KVO rate
STEP
delay
time
ramp up
time
ramp down
time
run time
V9
V10
V1
V2
V3
V4
V5
V6
V7
V8
Figure 5. Sample Ramp Program Delivery Profile (PDP)