Starting a Secondary Infusion
1.
Prepare secondary fluid
container and admini-
stration set.
a.
To perform automatic piggybacking,
a Continu-Flo
®
set from Baxter
must be used as the primary set.
b.
Follow directions accompanying the
products.
c.
Ensure all air is expelled from the
secondary set.
2.
Attach the secondary set
to the injection site of
the primary set above
the pump.
3.
Lower primary container
with the hanger accom-
panying the secondary
set.
4.
Set secondary flow rate
as follows:
a.
Press
SEC RATE
key. The right-hand
side of the main display now shows
secondary infusion data.
b.
Enter desired flow rate (in mL/hr)
for the secondary solution.
c.
To correct a mistake, press
CLR
or
SEC RATE
again and re-enter the cor-
rect value.
5.
Set secondary volume to
be infused as follows:
a.
Press the
SEC VTBI
key.
b.
Enter the desired fluid amount (in
mL) on the keyboard. Set VTBI
equal to the volume of fluid in the
container.
17