Operator's Manual

Flo-Gard® 6201

Volumetric Infusion Pump

Содержание Flo-Gard 6201

Страница 1: ...Operator s Manual Flo Gard 6201 Volumetric Infusion Pump...

Страница 2: ......

Страница 3: ...le the Pump Is Running 20 The OPTIONS Key 21 Using the Manual Programmed Delivery Profile 21 Using the Ramp Programmed Delivery Profile 24 Computer Control 29 Computer Monitoring 30 Locking the Front...

Страница 4: ...19 01 602 Original issue 4 92 Rev A 6 92 Rev B 4 93 Rev C 11 94 Rev D 3 99 Page Revision Front cover 0 i B ii D 1 B 2 11 0 12 B 13 14 0 15 38 B 39 C 40 46 B Back cover D A zero in the Revision column...

Страница 5: ...istration Safety clamp automatically occludes the tubing when the pump door is opened Occlusion sensors detect both upstream and down stream restrictions Sensitivity of the downstream occlusion sensor...

Страница 6: ...over to the primary program upon completion Volume Time programming automatically calcu lates flow rate Automatic self test routine checks for proper func tion before use Five hour memory retains inf...

Страница 7: ...ered when pressed The key is also used to advance to the next step when programming a delivery profile See NEXT legend item 4 above 6 CLEAR TOT VOL key Resets the total volume delivered to zero when t...

Страница 8: ...ART 7 8 9 4 5 6 1 2 3 0 CLR ALARM PRI VTBI PRI RATE ON OFF CHARGE STOP SEC RATE SEC VTBI SEC START OPTIONS NEXT COMPUTER CONTROL MONITOR P U S H VOLUMETRIC INFUSION PUMP Flo Gard 6201 1 2 3 4 5 6 7 8...

Страница 9: ...mming values and decimal point are entered with these keys 21 PRI VTBI key Allows programming of the primary VTBI 22 PRI RATE key Allows programming of the primary infusion rate 23 PRI START key Start...

Страница 10: ...Air sensor Detects air bubbles in the tubing The sensi tivity is adjustable via the pump s configura tion 5 Safety clamp Prevents accidental fluid flow by automati cally occluding the tubing whenever...

Страница 11: ...7 Figure 2 Pump Head Features 1 2 3 4 5 6 7...

Страница 12: ...only 5 VOLUME knob Adjusts volume of alert and alarm tones The alarm and alert tones cannot be turned com pletely off 6 PANEL LOCK switch Disables all front panel keys except BACK LIGHT and TOT VOL S...

Страница 13: ...9 PANEL LOCK VOLUME COMMUNICATIONS PORT 1 2 PUSH 3 4 5 6 7 8 9 Figure 3 Rear of Pump...

Страница 14: ...clinical applica tions Use only with standard Baxter soft tubing admini stration sets with an s designation Read and understand this manual before using the Flo Gard 6201 pump When infusing fluid thro...

Страница 15: ...ural delivery It is strongly recommended that pumps used for epidural drug delivery be clearly differentiated from pumps used for other routes of administration WARNING Epidural administration of drug...

Страница 16: ...serted Proceed to step c Note During an emergency in the alert mode the pump will operate without inserting the slide clamp An alert tone will sound until the slide clamp is inserted or the SILENCE k...

Страница 17: ...pump door and removing set 8 Attach set to I V access site 9 Turn pump on by press ing the ON OFF CHARGE key Verify that the pump performs the fol lowing self test a All segments of both displays app...

Страница 18: ...ty clamp Close the door and open the regulating clamp 10 Set VOLUME knob on the rear of the pump to the desired level Starting a Primary Infusion 1 Set primary flow rate as follows a Press PRI RATE b...

Страница 19: ...NG LED lights and a moving bar appears next to the flow rate setting b If the alarm software versions later than 1 08 or the alert tone sounds with the message INSERT SLIDE CLAMP when you press PRI ST...

Страница 20: ...lert tone and switches to a KVO Keep Vein Open rate of 5 mL hr or the current rate setting whichever is lower 8 Remove set from pump as follows a Press the STOP key b Close administration set regulati...

Страница 21: ...ove the pump 3 Lower primary container with the hanger accom panying the secondary set 4 Set secondary flow rate as follows a Press SEC RATE key The right hand side of the main display now shows secon...

Страница 22: ...program When this happens if the primary flow rate is set above 999 mL hr be sure to close the secondary set on off clamp Using Volume Time Programming Volume Time Programming allows you to select a V...

Страница 23: ...e PRI RATE or SEC RATE key a The pump calculates the flow rate required to deliver the desired VTBI in the specified time period b If the calculated flow rate is higher than the pump s capabilities th...

Страница 24: ...n ning NEW RATE appears in the message dis play and an alert tone sounds periodi cally 2 Enter the new flow rate on the keyboard a If the primary rate is changed above 999 mL hr ensure that the secon...

Страница 25: ...uration Note If the word OPTIONS is not visible when the pump is powered on you cannot use computer control or PDP Consult qualified hospital biomedical personnel re garding the appropriateness of the...

Страница 26: ...ps a If a previously set Program Delivery Profile is stored in the pump s mem ory STEP 1 appears in the message display and the rate and VTBI of the first programmed step appear in the main display b...

Страница 27: ...rt to enter a PDP 4 Modify previous Pro grammed Delivery steps a Press the NEXT key to reach the Pro gram Delivery step you wish to modify b Press the key for the infusion pa rameter you wish to modif...

Страница 28: ...e same manner After enter ing settings for the last step press NEXT again until the first step is dis played d To begin the Program Delivery Pro file press PRI START The pump will begin delivering flu...

Страница 29: ...ing this phase the pump delivers fluid at 5mL hr KVO rate Run Time This represents the total time of the Ramp PDP excluding the delay time Ramp Up Time This represents the time from the start of the t...

Страница 30: ...d the NEXT legend over the TOT VOL STATUS key is illuminated The pump is now in the PDP mode If PGM DELIV appears the pump is configured for manual Program De livery Consult with authorized hospital p...

Страница 31: ...to modify b Press the CLR key to erase the pre vious parameter c Enter the new value on the numeri cal keyboard d Press the NEXT key to lock in the pro grammed value and advance to the next parameter...

Страница 32: ...e NEXT key to lock in the pro grammed value and advance to RUN TIME h Press the CLR key to allow the entry of a new value i Enter the value on the numerical keyboard j Press the NEXT key to lock in th...

Страница 33: ...PDP appears in the message display if the pump is turned on within one hour of being turned off b Press PRI START to resume the Ramp PDP and the pump begins delivering fluid The pump will switch to t...

Страница 34: ...ont Panel The front panel can be locked during pumping to prevent tampering It can be unlocked at any time The TOT VOL STATUS and BACK LIGHT keys are not affected by the lock out This allows routine i...

Страница 35: ...CHECK display is a dashed line that stretches from left to right as the resistance to flow increases The left side of the display is labeled norm and corresponds to the pressure when the infusion was...

Страница 36: ...disonite Edison Chemical Co TOR or Hi Tor Plus Huntington Labs Bafix Hysan Corp A solution of 10 bleach and water Soapy water Isopropyl alcohol up to 95 Note Do not clean disinfect or sterilize any pa...

Страница 37: ...s not loaded into the slide clamp slot Open the pump door and insert the slide clamp into the slot Close the door NEW RATE No change until proce dure is com plete Flow rate is being changed while pump...

Страница 38: ...pabilities is calculated the message Hi or Lo will be displayed in the rate display If an attempt is made to start the pump with either of these messages displayed a FLOW RATE alert will be triggered...

Страница 39: ...e the Ramp PDP where it was interrupted Do one of the following press PRI START to start fluid delivery or press STOP to cancel the alert COM TIMEOUT No change There has been no communication between...

Страница 40: ...prevent the pump from restarting automatically press any enabled key such as SILENCE during the alarm INSERT SLIDE CLAMP Note This alarm option is available only on pumps running software versions lat...

Страница 41: ...e plies The condition has been caused by the computer not the pump To clear the alarm press OPTIONS to return the pump to computer control or press SILENCE or any pro gramming key to control the pump...

Страница 42: ...0 Hz Power Cord 2 9 m 9 5 ft long with Hospital Grade plug Fuse 0 5 A 250V SB 5 2 mm 13 64 in 20 mm 25 32 in Leakage Current Less than 50 microamps using UL 544 specified test methods Weight Approxima...

Страница 43: ...rsonnel Delay Time 00 00 08 00 hours in minute increments Run Time 01 00 24 00 hours in minute increments Ramp Up 00 00 to 45 of the Run Time in minute in crements Ramp Down 00 00 to the Run Time minu...

Страница 44: ...or SEC START key To view each subsequent option press the NEXT or SEC START key again c To stop viewing the configuration again press TIME and TOT VOL STATUS simultaneously Note If there is any quest...

Страница 45: ...e self test when the pump is turned on AUDIB SWI 1 OFF 2 ON OFF Number of Automatic Re starts The number of times the pump will restart auto matically after a down stream occlusion occurs and is then...

Страница 46: ...trol BAUD RATE 1 300 2 1200 3 2400 4 4800 5 9600 9600 Computer Control Selects the type of com puter control available at power up COMP CONT 1 DISABLED Computer Control not avail able 2 OFF W ALRM th...

Страница 47: ...lamp Loading Op tion Note The ALARM op tion is available only on pumps running software versions later than 1 08 INS CLAMP 1 OFF 2 ON for software versions earlier than 1 09 or ALERT for software vers...

Страница 48: ...saved indefinitely until changed from the front panel 4 PERM Profile is saved in permanent memory If a new PDP is entered the pump will revert to the PDP in perma nent memory once the cycle has been c...

Страница 49: ...y to the original purchaser and is not assignable or transferable and shall not apply to auxil iary equipment or disposable accessories THERE ARE NO OTHER WARRANTIES INCLUDING ANY IMPLIED WARRANTY AND...

Страница 50: ...tainer or in another Baxter ap proved container that will provide adequate protection dur ing shipment To ensure prompt return a Baxter Product Service representative must be notified before shipping...