PD-TaD 62
Connecting the system
822-187-3
49 / 114
No.
Cable
Step
On the laptop:
4
Ethernet cable
Use the supplied Ethernet cable to connect the laptop to the Power Box
(
Laptop
port).
–
–
If you do not want to run the laptop off its battery, connect the laptop to the
Power Box to establish a power supply. To do so, use the connection cable
with the C14G plug.
At the PD-TaD 62:
5
PoE cable
Connect the PoE cable:
to the Power Box (
PD Detector
port) and
to PD-TaD 62 (
Power Box
port)
This is used for power supply and data transmission at the same time.
6
tan δ cable
Connect the tan δ cable:
to the VLF HV generator (
PD Detector
port) and
to PD-TaD 62 (
VLF Generator
port).
7
–
Connect the HF filter to the
HV IN
port.
Further information: Chapter
Connecting the HF filter to the PD-TaD 62
(on
8
HV connection cable of
the VLF HV generator
Connect the HV connection cable to the HF filter at the
HV IN
port.
Further information: Chapter
Connecting the HF filter to the PD-TaD 62
(on
9
Short-circuit cable of
PD-TaD 62
On the PD-TaD 62, remove the short-circuit cable from ports
HV IN
and
HV OUT
.
At the far end of the test object:
10
–
Clean the cable termination of the phase to be tested carefully.
11
–
Attach the anti-corona hoods of the anti-corona protection device to the
cable termination of the phase to be tested.
12
Protective ring
(conductive hook and
loop fastener tape)
Attach a protective ring to the cable termination of the phase to be tested
–
directly over the screen.
Important:
Ensure that the protective ring does not come into contact with
the screen.