Maintenance
PD-TaD 62
100 / 114
822-187-3
18.3 Replacing the device protection fuses in the Power Box
Prerequisites
Flat-blade screwdriver, size 1.2 x 6.5 mm
Device protection fuses: 2 x T 16 AH (250 V / 16 A time lag), rated
breaking capacity 1500 A (H)
Procedure
1
Fuse cartridge
2
Fuse
3
Groove
1. Unscrew the fuse cartridge (1) out of the groove (3).
2. Replace the fuse (2).
3. Screw the fuse cartridge (1) back into the groove (3).