www.baumer.com
The CombiTemp™ TCR6 can be mounted in more different ways.
1.
Sensor tube without connection
Baumer offers compression glands fitting Ø6 and Ø8 mm sensor diameter.
This type of mounting is normally used for mounting a sensor direct into a non-pressurized application. If pressurized, ensure that the
connection is tightened correct, so no leakage occur.
A duct channel mounting flange for 8 mm sensor is also available (type: ZPX1-002)
All threaded connection can be mounted direct into the application without thermowell, however often a thermowell is required to enable the user
to take out the sensor for e.g. calibration without opening the system.
2.
Sensor with male threaded process connection G 1/2 A
This G 1/2 A connection can be mounted in a Baumer thermowell type ZPT4.
The process connection available are R 1/2, G 1/2 A, G 3/4 A, M20×1,5
or with hygienic ISO 2852 clamp DN 38
3.
Sensor with male threaded process connection G 3/4 A and G 1 A and sensors with G 1/2 or G 3/4 female thread
For these connections a special thermowell can be supplied.
Mount the gland/pocket into the application and install the sensor after the gland/pocket is fixed to the application. This will ensure that the cable is
not twisted during mounting.
Baumer recommend to use a thermal compound filled into the thermowell ensure best possible heat transfer between the pocket and the Com-
biTemp TCR6.
Baumer offers a 6 gram bag Thermal compound, type ZPX1-001.
Mounting
Electrical connection
Temperature transmitter
Ceramic terminal block
Pt100
2-w 4-w 3-w 2-w
- +
4...20 mA
Pt100
2-w
3-w
4-w 2-w
White
White
Red
Red
Warning
Note: Check the maximum temperature for the cable used.
Ambient temperature with built in transmitter: -40...85°C
Page 3 / 16
Operating instructions: 11119033 08
English
Operating instructions for temperature sensor
TCR6
Industrial temperature sensor
Ambient temperature with ceramic block: -40...160°C
Cable sensor insert
Single element, 4-wire
Single element, 2-wire
Double element
M12 plug version
Transmitter:
Pin 1 = 4...20 mA +
Pin 3 = 4...20 mA -
Ceramic block, 2-wire:
Pin 1/2 = 1st wire
Pin 3/4 = 2nd wire
Ceramic block, 4-wire:
Pin 1 = 1st wire
Pin 2 = 2nd wire
Pin 3 = 3rd wire
Pin 4 = 4th wire
Ceramic block, double element
Pin 1 = Element 1, 1st wire
Pin 2 = Element 1, 2nd wire
Pin 3 = Element 2, 1st wire
Pin 4 = Element 2, 2nd wire