14.
Process conditions
Process connection
BCID
Order
-
ing
key
Process
pressure
[bar]
Process
temperature,
standard
[°C]
Process tem-
perature, with
cooling neck
[°C]
Process tem-
perature, with
cooling neck
and spacer
Tamb = 20 °C Tamb = 20 °C Tamb = 60 °C
TFRN
Sleeve Ø6
T65
10
-1 … 40
-50 … 250
-50 … 400
-50 … 400
G 1/2 A DIN 3852-E
G51
11
-1 … 100 -50 … 250
-50 … 400
-50 … 400
G 1/2 A DIN 3852-A
G44 12
-1 … 100 -50 … 250
-50 … 400
-50 … 400
R 1/2 ISO 7/1
R01
13
-1 … 100 -50 … 250
-50 … 400
-50 … 400
1/2-14 NPT
N02
30
-1 … 100 -50 … 250
-50 … 400
-50 … 400
TFRH
G 1/2 A hygienic
A03
51
-1 … 100 -50 … 250
-50 … 400
N/A
BHC 3A DN 38
B01
60
-1 … 40
-50 … 250
-50 … 400
N/A
ISO 2852 DN38 (Tri-Clamp) C04
65
-1 … 40
-50 … 250
-50 … 400
N/A
ISO 2852 DN51 (Tri-Clamp) C05
66
-1 … 40
-50 … 250
-50 … 400
N/A
Varivent® Type N
V02
70
-1 … 16
-50 … 250
-50 … 400
N/A
Process temperature
85°C
0
50
100
150
200
250
300
350
400
20
30
40
50
60
70
80
90
100
Process
temperature
(°C)
Ambient temperature max. (°C)
Without cooling neck
With cooling neck
With cooling neck and spacer
(only TFRN)
16 /