GF Health Products, Inc. - www.grahamfield.com
Zenith 5000 Service Manual 999-0856-190C - DEC. 2013
7
ZENITH SERIES
WARNING!
DO NOT LOWER BED WHEN OBJECTS ARE
BENEATH BED. FAILURE TO INSPECT UNDER
BED CAN RESULT IN DAMAGE TO PROPERTY
OR PERSONAL INJURY.
ATTENTION:
S’assurer de ne pas faire descendre le lit
lorsque des objets se trouvent sous le lit. Ne
pas inspecter le dessous du lit pourrait entrainer
des dommages materiels et des risques de
blessures.
WARNING!
DO NOT LOWER BED WHEN
OBJECTS ARE BENEATH BED.
FAILURE TO INSPECT UNDER
BED CAN RESULT IN DAMAGE
TO PROPERTY OR PERSONAL
INJURY.
CAUTION
THIS BED IS SUITABLE FOR USE ONLY WITH
OXYGEN ADMINISTERING EQUIPMENT OF
THE NASAL OR MASK TYPE OR A TENT
COVERING ONLY THE UPPER HALF (HEAD
END) OF THE BED. OXYGEN TENT CANOPIES
SHOULD NOT EXTEND BELOW BED SPRING
LEVEL. LOCK HAND CONTROL AT FOOT OF
BED WHEN USING OXYGEN ADMINISTERING
EQUIPMENT.
ATTENTION:
CE LIT PEUT ETRE UTILISE UNIQUEMENT
AVEC UN EQUIPMENT DESTINE A
L’ADMINISTRATION D’OXYGENE DE TYPE
NASAL OU MASQUE OU AVEC UNE TENTE
RECOUVRANT SEULEMENT LA MOTTIE
AVENT (TETE) DU LIT. LES COTES DE LAS
TENTE OXYGENE NE DOIVENT PAS SE
PROLONGER PLUS DAS QUE LA SOMMIER
DU LIT.
The following warning labels have been placed on the bed to help protect you and the bed from damage.
Please do not remove any labels from the bed.
ENVIRONMENTAL SPECIFICATIONS AND STORAGE
AND RFI INFORMATION
Operating Conditions
Operation of the bed is based on the following conditions: Ambient Temperature of +10˚C to +40˚C; Relative Humidity of 30%
to 75% (Non-condensing); Atmospheric Pressure of 700hPa to 1060hPa; and a Splash Protection of IEC 60529.
Storage
Storage of the bed is based on the following conditions: Ambient Temperature of -10˚C to 70˚C; Relative Humidity of 10% to
100%; and an Atmospheric Pressure of 500hPa to 1060hPa.
Radio Frequency Interference (RFI)
RFI influences most electronic equipment. Caution should be exercised with regard to the use of portable communications
equipment in the area around such equipment. If RFI causes erratic behavior, shut the bed off immediately. Leave it off while
the transmission is in progress.
7