Zenith 5000 Service Manual 999-0856-190C - DEC. 2013 

4

ZENITH SERIES

This product is a variable height, adjustable mattress
platform, which will provide comfort and convenience
for residents/patients and caregivers in long term
care settings.

The MAXIMUM SAFE WORKING LOAD for the
Zenith 5000 Bed with weight evenly distributed, 
including bedding, resident/patient, support surface,
and all accessories, is 500 lbs. 

NEVER operate the bed if a Power Cord or Plug is
damaged or not working properly. Contact qualified
Service Personnel for examination and repair. Always
unplug the Power Cord when performing any mainte-
nance on the bed.

DO NOT open assemblies such as the Actuators, Hand
Control, or Control Box. If unauthorized personnel
perform work on these components, the manufacturer’s
warranty becomes void.

DO NOT use unauthorized parts, accessories, or
adaptors other than those specified/authorized by
GF Health Products, Inc.

When operating the HI/LO, Knee, or Back Functions
of the bed, ALWAYS ensure that the confined individual
is positioned properly within the confines of the bed.
DO NOT let any extremities protrude over the side or
between the bed rails when performing these functions.

DO NOT lower the bed when objects are beneath it.
Failure to inspect under the bed can result in damage
to property or personal injury.

The bed’s Pendant Cord MUST BE ROUTED AND
SECURED PROPERLY to ensure it does not become
entangled and eventually severed during use. Also
make sure all electrical cords DO NOT get tangled
around the bed, side rails, or legs during transport or
normal operation of the bed.

When using nasal-type or masked-type administering
equipment, all oxygen or air tubing MUST BE ROUTED
AND SECURED PROPERLY to ensure that the tubing
does not become entangled and eventually severed
during the normal operation of the bed.

Keep all moving parts free of obstructions (i.e. blankets/
sheets, heating blankets/pads, wiring, etc.).

Beds manufactured by Basic American Medical Products
are designed for use within an institutional healthcare 
environment (i.e. assisted living, skilled nursing, tran-
sitional care, rehabilitational care, etc.).

DO NOT use the assist devices as push handles for moving
the bed. Assist devices can be deformed or broken if
excessive side pressure is exerted. Assist devices are not
meant for patients considered as high risks for entrapment
(i.e. patients with pre-existing conditions such as 
confusion, restlessness, lack of muscle control, altered
mental status, either organic or medicinal, or a combi-
nation thereof). Additional safety measures should be
considered for such high-risk patients. 

NEVER permit more than one (1) person on/in the bed
at any time.

Body weight should be evenly distributed over the
sleeping surface of the bed. DO NOT allow the patient
to lay, sit, or lean in such a way that their entire body
weight is placed only on the raised head or foot sections
of the bed. This especially applies when repositioning
or transferring a patient in or out of the bed. Increased
risk may occur when the patient’s size and/or weight are
inappropriate for the bed’s dimensions or weight capacity.

Risk of entanglement or injury may occur if the mattress
used with mattress retainers does not fill the entire width
between stops or which compresses to less than 1.50
inches under user’s weight.

Mattress must be properly sized to fit the mattress support
platform and must remain centered on the support platform
relative to State and Federal guidelines. Recommended
minimum dimensions of mattress is 35 inches wide and
6 inches deep. Length should match the mattress support
platform. Use of an improperly fitted mattress could 
result in injury or death.

IMPORTANT:

Powered air mattress surfaces may pose

a risk of entrapment. Prior to use, ensure the therapeutic
benefits outweigh the risk of entrapment.

The bed is intended for use, storage, and transport within
a temperature range of -40˚C to +60˚C. It has a water-
resistance rating of IPX6 and IS NOT to be power-
washed or submersed. 

!

!

!

!

!

!

!

!

!

!

!

!

!

!

!

!

!

!

IMPORTANT SAFETY AND WARNING INFORMATION

Содержание Zenith 5000

Страница 1: ...l sections pertaining to the bed model before use GF Health Products Inc www grahamfield com Zenith Series Service Manual 999 0856 190C DEC 2013 ETL UL CSA APPROVED UL 60601 1 IEC 60601 2 38 CSA 22 2...

Страница 2: ...he time of printing The most current product information can be found online at www grahamfield com Please check all parts for shipping damage and test before using In case of damage DO NOT USE Basic...

Страница 3: ...Wallsaver Installation 12 Plugging in Your Optional Staff Nurse Control 13 Pendant Hand Controller Operation 14 Optional Staff Nurse Control Operation 15 Stationary Low Roll Positioning 16 SECTION B A...

Страница 4: ...ng blankets pads wiring etc Beds manufactured by BasicAmerican Medical Products are designed for use within an institutional healthcare environment i e assisted living skilled nursing tran sitional ca...

Страница 5: ...bed components acces sories that correspond with the needs of the patient resident and the particular bed you have purchased Matching the correct bed component that correlates with the regulatory gui...

Страница 6: ...ottom wheels on both the head end and foot end for any damage or wear Replace if needed X Check all rolling casters to ensure that they roll properly and are unobstructed X Electrical Inspection of Co...

Страница 7: ...INE A L ADMINISTRATION D OXYGENE DE TYPE NASAL OU MASQUE OU AVEC UNE TENTE RECOUVRANT SEULEMENT LA MOTTIE AVENT TETE DU LIT LES COTES DE LAS TENTE OXYGENE NE DOIVENT PAS SE PROLONGER PLUS DAS QUE LA S...

Страница 8: ...GF Health Products Inc www grahamfield com Zenith 5000 Service Manual 999 0856 190C DEC 2013 ZENITH SERIES ZENITH 5000 Bed Set up Instructions SECTION A 8...

Страница 9: ...ox DISCARD 1 Cable Tie on Bundle 2 Large Foam Block 3 Large Cable Tie CUT 4 Notched 2 x 4 Board 5 Foam Wrapping 6 End Caps with Foam 7 Small Cable Tie CUT 8 Notched Leg Foam KEEP 9 Wireform Wallsaver...

Страница 10: ...sider this the outside of the board Position 2 mounting tubes on the inside of the footboard with L facing outward Align the top hole of the mounting tubes with the top holes in the footboard Insert a...

Страница 11: ...76 OR 80 CONFIGURATIONS SEE PAGES 10 AND 12 1 Determine if you need to position your bed in an 80 or 76 configuration 80 outside holes 76 inside holes see Closeups A B 2 On the foot end carefully sque...

Страница 12: ...he caster base slots WALLSAVER POSITIONS 1 The Zenith 5000 wallsaver features two standard positions for easy bed mattress length reconfigurations 76 and 80 2 For 76 beds position the wallsaver ends i...

Страница 13: ...IL A CLOSEUP OF STAFF CONTROL CABLE CONNECTION DETAIL B MAKE SURE LOCK END CAPS ARE SCREWED ON SECURELY SEE PAGE 10 OF THIS MANUAL FOR PROPER ASSEMBLY OF ELBOW MOUNTING BRACKETS TUBES FOOTBOARD Female...

Страница 14: ...CONTROL BOX AND PLUGS INTO A SMALL BLACK CONTROL CABLE WHICH PLUGS DIRECTLY INTO THE CONTROL BOX IN ORDER TO MAXIMIZE THE EASE OF USE AS WELL AS THE LONGEVITY OF THE ZENITH BED THE MICRO PROCESSOR CO...

Страница 15: ...2 To lower the head back deck to horizontal press Select 1 3 3 To lockout the head back pendant function press Select 1 4 Lockout light 5 will glow 4 To unlock the head back pendant function press Sel...

Страница 16: ...sy mobility BED OPERATIONS STATIONARY LOW ROLL POSITIONING STATIONARY POSITION NO ROLL Raise the Bed Up Horizontally to Desired Height Press the HI LO Up Button MOBILE POSITION LOW ROLL Lower the Bed...

Страница 17: ...GF Health Products Inc www grahamfield com Zenith 5000 Service Manual 999 0856 190C DEC 2013 17 ZENITH SERIES ASSIST DEVICES Model ZG58174 SECTION B...

Страница 18: ...e 2 To set the assist bar back to its upright locked position grab the top of the assist bar with one hand and pivot the assembly upward until the black knob mechanism snaps into place locking the ass...

Страница 19: ...IL FOUR LANYARD ASSEMBLIES AND 2 PHILLIPS SCREWS OR 2 OPTIONAL HEX HEAD SCREWS ORDER BAG 999 0842 905 FOR JUST THE REPLACEMENT HARDWARE THE PLASTIC RAIL CAN BE POSITIONED ON EITHER THE HEAD OR FOOT DE...

Страница 20: ...bag Do not remove the 4 four 1 4 20 inserts 3 from the board They will be used again to mount the board to the trapeze support 3 Assemble the Trapeze Support 1 to the main frame tubes using the two Cl...

Страница 21: ...GF Health Products Inc www grahamfield com Zenith 5000 Service Manual 999 0856 190C DEC 2013 21 ZENITH SERIES Mechanical Electrical Information SECTION C...

Страница 22: ...floor surface to top of sleep deck NOTE All dimensions are in a range of 25 inches GF HEALTH PRODUCTS INC DUTY CYCLE 2 min Continuous 18 min MODE OF OPERATION Intermittent 500 lbs SAFE WORKING LOAD I...

Страница 23: ...e on the bed Pendant Holster Part 999 0791 000 Quantity 1 included with your bed Clipped on Pendant when bed first shipped Control Junction Box Part 999 0844 300B Quantity 1 Box and cable color White...

Страница 24: ...HEAD CABLE FOOT CABLE HI LO CABLE FOOT HI LO CABLE HEAD BLACK CONTROL CABLE GROUND WIRE u v w x y z u u u u v v v v w w y y x x z z ZENITH 5000 SERIES ELECTRICAL CABLING Power Cable Plug in Optional B...

Страница 25: ...GF Health Products Inc www grahamfield com Zenith 5000 Service Manual 999 0856 190C DEC 2013 25 ZENITH SERIES SERVICE REPLACEMENT PARTS SECTION D...

Страница 26: ...ice 999 0843 251SP HI LO Motor Cable attached Hardware Pack 1 Motor 999 0844 052SP OR Head Motor Cable attached Hardware Pack 1 Motor 999 0844 053SP OR Foot Motor Cable attached Hardware Pack 1 Motor...

Страница 27: ...rvice Pack INCLUDES 1 Pkg PLEASE SEE PAGE 13 FOR INSTALLATION INSTRUCTIONS 1 Switch Pad with Bezel Cables 1 Shroud Cable Cover 6 6 Phillips Head Truss Screws 3 4 Nylon Cable Tie 1 Installation Instruc...

Страница 28: ...ing hardware installation instruction sheet 4 TO ORDER A REPLACEMENT CONTROL BOX PLEASE USE 999 0844 300B Includes control box detachable power cord control cable ground wire screw cable ties installa...

Страница 29: ...3 Unplug the existing battery and discard Plug in the new battery to the Connector Cable and reinsert into slot Replace battery cover NOTE THE 9V BATTERY INCLUDED WITH YOUR CONTROL BOX IS ONLY GOOD FO...

Страница 30: ...GF Health Products Inc www grahamfield com Zenith 5000 Service Manual 999 0856 190C DEC 2013 30 ZENITH SERIES FRAMING SERVICE PARTS For All Zenith 5000 Bed Models SECTION E...

Страница 31: ...ng hardware installation instruction sheet 2 TO ORDER A REPLACEMENT KNEE DECK USE SERVICE PACK 999 0844 940SP includes deck mounting hardware installation instruction sheet 3 TO ORDER A REPLACEMENT FO...

Страница 32: ...856 190C DEC 2013 32 ZENITH SERIES SERVICE REPLACEMENT PART DIAGRAM MAIN FRAME CAPPING CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS...

Страница 33: ...ND FOOT DECKS CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS PLUGS ON THE BED SEE BOTTOM OF PAGE 27 FOR ORDER NUMBERS CAPPING ON ARMS...

Страница 34: ...GF Health Products Inc www grahamfield com Zenith 5000 Service Manual 999 0856 190C DEC 2013 34 ZENITH SERIES TROUBLESHOOTING All ZENITH 5000 Models SECTION E...

Страница 35: ...t c If Pendant works from the Control Box replace the Control Cable Please refer to page 27 in this Service Manual for replacement part STAFF CONTROL NOT WORKING LIGHTS BLINKING 1 Check all Staff Cont...

Страница 36: ...eck Motor Condition Cracked Housing Scratches Bent Shaft etc 3 Is Function Locked Out on Staff Control 4 Head Foot Motor Works But HI LO Does Not a Unplug Power Cord b Remove Cord Guard at Control Box...

Страница 37: ...the cutout on the outside of the footboard h Secure the Staff Control by re inserting the two Phillips Screws 2 you removed in Step e i Re position the Shroud on the inside of the foot board and secur...

Страница 38: ...__________________________________ _________________________________________________________________________________ _________________________________________________________________________________ _...

Страница 39: ...se price Any repair work or replacement shall not exceed the warranty duration Basic American Medical Products will at its option repair or replace defective product or parts provided that the replace...

Страница 40: ...Lac WI 54937 Customer Service 1 800 365 2338 Fax 920 929 8213 www grahamfield com 2013 GF Health Products Inc All Rights Reserved Basic American Medical Products and Zenith are registered trademarks...

Отзывы: