ZENITH SERIES
WALLSAVER ASSEMBLY
1. Position the Wireform Wallsaver
(Part 999-0856-180)
with bent end facing upward and tab ends facing inward
as shown at right.
2. Determine the position desired (See bottom of page).
3. Gently squeeze the tab ends of the wallsaver
inward toward the center of the wallsaver and,
holding the tabs parallel with the slots in the
caster bases (See DETAIL A), slide the tabs
into the slots while letting the wallsaver gently
expand outward.
4. Turn the wallsaver downward until it rests on
the floor (See Detail B).
WALLSAVER REMOVAL
1. To remove or move the Wireform
Wallsaver to a new position, raise
the wallsaver off the floor until
the end tabs are horizontal.
2. Squeeze the ends toward the
center of the wallsaver until the
end tabs slide out of the caster
base slots.
WALLSAVER POSITIONS
1. The Zenith 5000 wallsaver features two
standard positions for easy bed/mattress
length reconfigurations - 76” and 80”.
2. For 76” beds, position the wallsaver ends
in the BACK holes in the caster base slots
(toward the foot end). For 76” beds using
an optional Trapeze unit, position the wall-
saver ends in the FRONT holes in the caster
bases (toward the head end.)
3. For 80” beds, position the wallsaver ends
in the FRONT slots in the caster bases
(toward the head end of the bed).
IMPORTANT NOTE: WHEN POSITIONING THE BED’S
WALLSAVER FOR A PARTICULAR BED LENGTH
(I.E. 76” OR 80”), MAKE SURE THAT THE OTHER
BED ADJUSTMENTS (HEADBOARD AND FOOT-
BOARD POSITIONS AND MATTRESS RETAINER
POSITION) ARE IN THE SAME CONFIGURATION.
PLEASE REFER TO PAGES 10 AND 12.
Wireform Wallsaver
999-0856-180
Bend faces upward
Start with tab
held horizontally
Finish with Wallsaver
resting on floor
Use Back hole
for 76” position
Use Front hole
for 80” position
Brace cut
away for
visual clarity
Brace cut
away for
visual clarity
STANDARD WIREFORM WALLSAVER INSTALLATION