5. Configuration Menu
The Configuration Menu is shown in Figure 4-9 above. Below is a description of each menu item:
Network Configuration:
Used to set the IP address and other network parameters of the
Industrial Mux. This selection is only used when configuring the Mux for use on your network
rather than connecting directly to a pc.
o
If used on a DHCP enabled network and Option 1 is “No”, the Mux will obtain an
IP address automatically when connected to the network.
Figure 5-1:
Network Configuration sub-menu
Pyrometer Supply Voltage (mV):
This is the voltage that supplies power to each Exactus®
pyrometer.
(Only adjust after consulting with factory and if the pyrometer
communication cable is longer than 15meters; otherwise; it should remain at default
value of 6200mv.)
Mode Options:
Used to configure the alarms for individual pyrometers. More detail is
provided in Alarm Configuration section.
Mux Ambient Alarm Configuration:
Brings up the Mux Internal Temperature Configuration
Menus; this screen allows a user to define the allowable operating temperatures for industrial
Mux and how the Mux will react to over-temp conditions.
o
Internal Temp Upper Limit:
Temperature which will trigger error condition.
o
Internal Temp Error Closes Relay 1:
Will over-temp condition close Relay1?
o
Internal Temp Error Closes Relay 2:
Will over-temp condition close Relay2?
o
Relay polarity:
Unlike the other alarms, the polarity of the relay output is
selectable for the Mux over-temp condition. The reason is that if the Mux
overheats, it may become unable to keep a contact closed. By allowing for
selectable polarity, the user can choose to keep the relay in its activated state during
Содержание EXACTUS IFM8-I
Страница 4: ...3 Figure 1 2 IFM8 IA Overview...
Страница 10: ...9 c The following screen Figure 4 3 will appear Click the Properties button Figure 4 3...
Страница 15: ...14 Figure 4 10 Check the Echo typed characters locally check box and close the dialog windows Figure 4 11...
Страница 18: ...17 Figure 6 1 Alarm configuration menu Figure 6 2 Pyrometer Internal Temperature Alarm Configuration Menu...