EXACTUS IFD-5 and IFA-5 User Guide

   

 

3.9  Calibration 

3.9.1  Checking Calibration of the IFA-5 Analog Output Module 

Calibration of the analog output is very easy and straightforward.  A menu within TemperaSure 
provides an easy-to-use interface for the calibration of the analog output.  Checking the 
calibration does not modify the calibration of the IFA-5 Module.  To check the calibration, 
follow the steps below.  Instructions for changing the calibration follow this section.  Calibration 
of the analog output is not required before the first year of operation and only annually after the 
first year.   

To check the calibration of the IFA-5 Analog Output Module: 

1.

 

Make sure that the IFA-5 Module and the pyrometer are disconnected from any process, 
system, or tool that could be affected by readings from the IFA-5 Module or the pyrometer 
itself.   

2.

 

Verify that the voltage or current meter being used to calibrate the IFA-5 Module is properly 
calibrated and its calibration is up to date. 

3.

 

Connect the pyrometer to a computer capable of running the TemperaSure software under the 
Windows

 operating system.  Open the TemperaSure software and verify readings are 

coming from the pyrometer to the computer. 

4.

 

Connect the meter to the IFA-5 Module.  If the IFA-5 Module is to be used in voltage mode, 
the meter should be connected in parallel with the output.  If the IFA-5 Module is to be used 
in current mode, the meter should be connected in series.   

5.

 

Configure the IFA-5 Module for Internal Power (see instructions for Optional External 
Power).   

6.

 

Make sure the IFA-5 Module has been running for at least 30 minutes so it is fully warmed 
up.   

7.

 

Determine the COM port to which the pyrometer is connected.  To determine the COM port, 
click on the numbered box above the temperature readout for the pyrometer.  The COM port 
will be displayed in the Pyrometer section on the General tab. 

8.

 

Under the Options menu in TemperaSure, select Configure IFA-5 Module. A calibration 
menu box will pop up (see Figure 5). 

9.

 

Using the menu, select the COM port of the pyrometer that is attached to the IFA-5 Module 
and whether the IFA-5 Module is being used in voltage or current mode.  

10.

 

Measure the voltage or current from the IFA-5 Module.  For a voltage output there should be 
a potential of 10 volts across the +V out and –V out terminals.  For a current output there 
should be a 20mA current flowing between the +I out and –I out terminals (ExtPWR and –V 
out if External Power is utilized – see section on Optional External Power). 

3.9.2  Adjusting Calibration of the IFA-5 Analog Output Module  

To adjust the calibration of the IFA-5 Module, follow the steps outlined in the previous section.  
Adjustments are made by the up and down arrows on the calibration menu box.  Adjust the 
calibration so the output from the module is either 10.000 volts or 20.000mA, depending on the 
mode selected.  Once the output is correct, press the “Save” button on the calibration menu box.  
None of the adjustments will be stored until the “Save” button is pressed. 

 

Содержание EXACTUS IFA-5

Страница 1: ...ta proportions weights etc given herein may change without prior information and do not constitute the agreed contractual quality of the product It is the responsibility of the recipient of our produc...

Страница 2: ...Optics either a lightpipe or lens assembly Ex 100 Electronics SST tube dimension 1 0 diameter 4 0 long 25 4mm 102mm IFD 5 Interface module for conditioning power and communicating with a PC Dimension...

Страница 3: ...em is powered with a DC Power Supply through the IFD 5 Digital Interface Module as shown in Fig 1 0 The IFD 5 Module may be powered by any DC voltage between 12 and 24VDC The IFD 5 supplies power for...

Страница 4: ...vides a connector designed for use when the two modules are stacked When the two modules are stacked vertically the power input and digital communications connections are pin for pin directly above on...

Страница 5: ...dule is supplied configured for internal power To configure for external power a jumper inside the module must be moved The IFA 5 Module and the IFD 5 Module are similar in appearance so make sure you...

Страница 6: ...es above the graphical display Open the Pyrometer Settings menu for the pyrometer that is to be configured and follow the steps below 1 Select the Analog Output tab See Figure 5 below 2 Click the Use...

Страница 7: ...e table If the temperature goes below the temperature assigned to 1 volt the output will stay at 1 volt If the temperature exceeds the temperature assigned to 5 volts the output will stay at 5 volts T...

Страница 8: ...atchdog Failure Alarm Timeout In ____ box The Configure IFA 5 Module menu accessed through Menu Options Configure IFA 5 Module is used to set the Watchdog Failure Signal Level current or voltage level...

Страница 9: ...the meter should be connected in series 5 Configure the IFA 5 Module for Internal Power see instructions for Optional External Power 6 Make sure the IFA 5 Module has been running for at least 30 minu...

Страница 10: ...to 30VDC Power Consumption with IFD 5 IFA 5 and Pyrometer sensor 3 watts 125mA 24VDC Pin 1 10 Assignments Pin 1 GND Pins 1 3 5 and 10 are internally connected Pin 2 Isolates pyrometer if open COM pow...

Страница 11: ...racy 0 02 of FSR Resolution 0 002 of FSR Readback Accuracy 1 of FSR Zero Drift Voltage output 10 V C Current output 0 2 A C Span Temperature Coefficient 5ppm C Accuracy 0 02 of FSR Resolution 0 002 of...

Страница 12: ...and IFA 5 User Guide 9 5 1 IFD 5 and IFA 5 Support Cabinets Figure 8 Electronics support cabinet layout example Integrated Instrument System Figure 7 Electronics support cabinet layout example Fiber...

Страница 13: ...tput C Cooling Fan if used Active air cooling from the air purge system is the preferred method of cooling the cabinet as it reduces the risk of introducing contaminants to the cabinet D Digital Inter...

Страница 14: ...st temperature The temperature should not exceed 70 C Exceeding this temperature will not damage the electronics but may affect the temperature measurement accuracy If the temperature exceeds 80 C fai...

Отзывы: