EXACTUS IFD-5 and IFA-5 User Guide

   

 

3.3  Wiring for External Power  

The wiring for the current mode is also different when the output is powered externally.  When 
an external 24VDC power supply is used, wire as shown in Figure 4.  

 

Figure 4

 

3.4  Software Setup 

Analog output is configured with TemperaSure under each pyrometer’s setup.  The analog output 
from each unit (channel) can be configured as a voltage or a current, 0 to 10 volts or 4 to 20mA.  
The minimum and maximum electrical values are entered, along with the corresponding 
minimum and maximum temperature values.  These four numbers will scale the analog output. 
See Figure 5: Analog Output Setup display. 

 

 

3.5  Configuration  

With all the connections made, open the TemperaSure software program.  Wait a few seconds 
for the instrument to start displaying temperatures in the box or boxes above the graphical 
display.  Open the Pyrometer Settings menu for the pyrometer that is to be configured and follow 
the steps below: 

1.

 

Select the Analog Output tab.  See Figure 5 below. 

2.

 

Click the “Use Analog Output” checkbox. 

3.

 

Click on either Voltage or Current button. 

4.

 

Go to the corresponding section and type in the minimum temperature and the maximum 
temperature and the corresponding voltage or current values.  Voltages between 0 and 10 
volts and currents between 0 and 20mA may be entered.  See Scaling section below for 
details. 

5.

 

Values may be stored in two ways.  By pressing the “Save” button the values are stored 
in the electronics memory and are retained even if the pyrometer is reset or powered 
down.  Values may also be stored by pressing the “Apply” button; however these values 
will not be retained if the pyrometer is reset or powered down.  Pressing the “Apply” 
button will interrupt the output from the pyrometer for two to four seconds.  Pressing the 
“Save” button will interrupt output from the pyrometer for several seconds.  

Содержание EXACTUS IFA-5

Страница 1: ...ta proportions weights etc given herein may change without prior information and do not constitute the agreed contractual quality of the product It is the responsibility of the recipient of our produc...

Страница 2: ...Optics either a lightpipe or lens assembly Ex 100 Electronics SST tube dimension 1 0 diameter 4 0 long 25 4mm 102mm IFD 5 Interface module for conditioning power and communicating with a PC Dimension...

Страница 3: ...em is powered with a DC Power Supply through the IFD 5 Digital Interface Module as shown in Fig 1 0 The IFD 5 Module may be powered by any DC voltage between 12 and 24VDC The IFD 5 supplies power for...

Страница 4: ...vides a connector designed for use when the two modules are stacked When the two modules are stacked vertically the power input and digital communications connections are pin for pin directly above on...

Страница 5: ...dule is supplied configured for internal power To configure for external power a jumper inside the module must be moved The IFA 5 Module and the IFD 5 Module are similar in appearance so make sure you...

Страница 6: ...es above the graphical display Open the Pyrometer Settings menu for the pyrometer that is to be configured and follow the steps below 1 Select the Analog Output tab See Figure 5 below 2 Click the Use...

Страница 7: ...e table If the temperature goes below the temperature assigned to 1 volt the output will stay at 1 volt If the temperature exceeds the temperature assigned to 5 volts the output will stay at 5 volts T...

Страница 8: ...atchdog Failure Alarm Timeout In ____ box The Configure IFA 5 Module menu accessed through Menu Options Configure IFA 5 Module is used to set the Watchdog Failure Signal Level current or voltage level...

Страница 9: ...the meter should be connected in series 5 Configure the IFA 5 Module for Internal Power see instructions for Optional External Power 6 Make sure the IFA 5 Module has been running for at least 30 minu...

Страница 10: ...to 30VDC Power Consumption with IFD 5 IFA 5 and Pyrometer sensor 3 watts 125mA 24VDC Pin 1 10 Assignments Pin 1 GND Pins 1 3 5 and 10 are internally connected Pin 2 Isolates pyrometer if open COM pow...

Страница 11: ...racy 0 02 of FSR Resolution 0 002 of FSR Readback Accuracy 1 of FSR Zero Drift Voltage output 10 V C Current output 0 2 A C Span Temperature Coefficient 5ppm C Accuracy 0 02 of FSR Resolution 0 002 of...

Страница 12: ...and IFA 5 User Guide 9 5 1 IFD 5 and IFA 5 Support Cabinets Figure 8 Electronics support cabinet layout example Integrated Instrument System Figure 7 Electronics support cabinet layout example Fiber...

Страница 13: ...tput C Cooling Fan if used Active air cooling from the air purge system is the preferred method of cooling the cabinet as it reduces the risk of introducing contaminants to the cabinet D Digital Inter...

Страница 14: ...st temperature The temperature should not exceed 70 C Exceeding this temperature will not damage the electronics but may affect the temperature measurement accuracy If the temperature exceeds 80 C fai...

Отзывы: