EXACTUS IFD-5 and IFA-5 User Guide

   

 

1.1  Digital and Analog Interface Modules 

The IFD-5 Digital Interface Module allows the pyrometer to be connected to any PC.  This 
module also isolates the electronics from any electrical noise and supplies a regulated power 
supply to the pyrometer. There are three connections on the IFD-5 Module: on one side of the 
module is a RJ45 connector for the pyrometer cable and a female DB9 connector for the host 
computer.  On the opposite side of the module is the connection for the power supply.  The 
interface module has a green LED that indicates when power is being applied.   

The pyrometer can communicate with a host computer using either RS-232 or RS-422 serial 
communication.  If it is configured for RS-232 than the interface module should be connected to 
the  PC or digital data logger with a cable no longer than 15 meters. 

BASF also offers an analog output option, the EXACTUS IFA-5, which can be configured to 
either a voltage or current output.  The value of the analog output can be scaled as needed using 
the TemperaSure

®

 

software from 0 to 10 volts or 0 to 20 mA for the desired temperature range.   

1.2  Power Supply 

A single channel system is powered with a DC Power Supply through the IFD-5 Digital Interface 
Module as shown in Fig 1.0.  The IFD-5 Module may be powered by any DC voltage between 12 
and 24VDC.  The IFD-5 supplies power for the electronics and optional IFA-5 analog module. 

1.3  Software Interface 

EXACTUS TemperaSure

®

 graphical user software gives the user the ability to log and/or display 

temperature readings and configure the pyrometer and IFA-5 module user settings.  Refer to the 
EXACTUS Optical Thermometer user manual on how to install and set up the software. 

2.0  IFD-5 Digital Interface Module 

Using the IFD-5 interface box, the EXACTUS pyrometer can communicate digitally, selectable 
for either RS232 or RS422 line levels.  The IFD-5 provides power to the pyrometer and isolation 
of the communication lines.  The Modbus and EXACTUS protocols are available using this 
interface. 

 

To connect the pyrometer to a standard PC serial port using the IFD-5, make sure that the 
selector switch for RS232/RS422 is in the RS232 position.  This is a small single DIP switch 
near the IFD-5’s DB-9 connector. 

IFD-5 Digital Pinouts 

Pin 

RS232 

RS422 

Unused 

TX+ 

RXD 

Unused 

TXD 

Unused 

Unused 

RX- 

Signal Gnd 

Signal Gnd 

Unused 

TX- 

RTS 

RTS * 

Unused 

RX+ 

Unused 

Unused 

 
 
 
 
Table 1:

   

RS232/RS422 Serial Connections for IFD-5  
 

*

 NOTE:  

For RS232 the RTS line must be 

held in an asserted (HIGH) state or be left 
unconnected. 

Содержание EXACTUS IFA-5

Страница 1: ...ta proportions weights etc given herein may change without prior information and do not constitute the agreed contractual quality of the product It is the responsibility of the recipient of our produc...

Страница 2: ...Optics either a lightpipe or lens assembly Ex 100 Electronics SST tube dimension 1 0 diameter 4 0 long 25 4mm 102mm IFD 5 Interface module for conditioning power and communicating with a PC Dimension...

Страница 3: ...em is powered with a DC Power Supply through the IFD 5 Digital Interface Module as shown in Fig 1 0 The IFD 5 Module may be powered by any DC voltage between 12 and 24VDC The IFD 5 supplies power for...

Страница 4: ...vides a connector designed for use when the two modules are stacked When the two modules are stacked vertically the power input and digital communications connections are pin for pin directly above on...

Страница 5: ...dule is supplied configured for internal power To configure for external power a jumper inside the module must be moved The IFA 5 Module and the IFD 5 Module are similar in appearance so make sure you...

Страница 6: ...es above the graphical display Open the Pyrometer Settings menu for the pyrometer that is to be configured and follow the steps below 1 Select the Analog Output tab See Figure 5 below 2 Click the Use...

Страница 7: ...e table If the temperature goes below the temperature assigned to 1 volt the output will stay at 1 volt If the temperature exceeds the temperature assigned to 5 volts the output will stay at 5 volts T...

Страница 8: ...atchdog Failure Alarm Timeout In ____ box The Configure IFA 5 Module menu accessed through Menu Options Configure IFA 5 Module is used to set the Watchdog Failure Signal Level current or voltage level...

Страница 9: ...the meter should be connected in series 5 Configure the IFA 5 Module for Internal Power see instructions for Optional External Power 6 Make sure the IFA 5 Module has been running for at least 30 minu...

Страница 10: ...to 30VDC Power Consumption with IFD 5 IFA 5 and Pyrometer sensor 3 watts 125mA 24VDC Pin 1 10 Assignments Pin 1 GND Pins 1 3 5 and 10 are internally connected Pin 2 Isolates pyrometer if open COM pow...

Страница 11: ...racy 0 02 of FSR Resolution 0 002 of FSR Readback Accuracy 1 of FSR Zero Drift Voltage output 10 V C Current output 0 2 A C Span Temperature Coefficient 5ppm C Accuracy 0 02 of FSR Resolution 0 002 of...

Страница 12: ...and IFA 5 User Guide 9 5 1 IFD 5 and IFA 5 Support Cabinets Figure 8 Electronics support cabinet layout example Integrated Instrument System Figure 7 Electronics support cabinet layout example Fiber...

Страница 13: ...tput C Cooling Fan if used Active air cooling from the air purge system is the preferred method of cooling the cabinet as it reduces the risk of introducing contaminants to the cabinet D Digital Inter...

Страница 14: ...st temperature The temperature should not exceed 70 C Exceeding this temperature will not damage the electronics but may affect the temperature measurement accuracy If the temperature exceeds 80 C fai...

Отзывы: