BASF EXACTUS IFA-5 Скачать руководство пользователя страница 11

EXACTUS IFD-5 and IFA-5 User Guide

   

 

 
 
 

IFA-5 Input/Output 

Input Voltage Range 

  (supplied by IFD-5)

 

Standard: 24VDC (+10VDC to +30VDC) 

Power Consumption (with IFD-5, IFA-5,   
and Pyrometer sensor) 

3 watts, 125mA, 24VDC 

Pin 1 – 10 Assignments 
 
(Isolated Analog Output - 3000VDC) 

Pin 1  +I (0  to 20mA output) 
Pin 2 

 

-I (0  to 20mA output 

Pin 3  +V (0 to 10VDC output) 
Pin 4  -V (0 to 10VDC output) 
Pin 5  Ext. Pwr (0 to 20mA loop power 24VDC) 
Pin 6  Init (N/C - for manufacture use only) 
Pin 7  D+ RS485 communications to IFD-5 
Pin 8  D- RS485 communications to IFD-5 
Pin 9  +Vs (power from IFD-5) 
Pin 10 GND (circuit ground) 

10 pin Terminal – pin spacing 

5.08mm 

Analog Outputs 

0 to 20mA 

0 to 10 Volts 

0 – 10 Voltage Output 

10mA maximum 

Accuracy 

± 0.02% of FSR 

Resolution 

± 0.002% of FSR 

Readback Accuracy 

± 1% of FSR 

Zero Drift 

Voltage output: ± 10µV/°C 

Current output: ± 0.2µA/°C 

Span Temperature Coefficient 

± 5ppm/°C 

Accuracy 

± 0.02% of FSR 

Resolution 

± 0.002% of FSR 

Current Load Resistance 

Internal power : 500 ohms 

External 24VDC power: 1050 ohms 

Update Rate 

Adjustable up to 100 readings per second. 

 

 

5.0  Mounting 

 

 

The backs of both the IFD-5 Digital Interface Module and IFA-5 Analog Output Module 
have attachments for easy DIN rail mounting.  To attach the modules to a DIN rail pull the 
orange tab at the bottom of the module out, slip the module onto the DIN rail and push the 
tab back in to secure the module to the rail.  A short piece of DIN rail is provided for 
mounting the Module to solid surfaces. 

 

It is recommended that the modules be mounted in a cabinet or NEMA enclosure rated for 
area of use.  See mounting examples in section 5.1 

 

Содержание EXACTUS IFA-5

Страница 1: ...ta proportions weights etc given herein may change without prior information and do not constitute the agreed contractual quality of the product It is the responsibility of the recipient of our produc...

Страница 2: ...Optics either a lightpipe or lens assembly Ex 100 Electronics SST tube dimension 1 0 diameter 4 0 long 25 4mm 102mm IFD 5 Interface module for conditioning power and communicating with a PC Dimension...

Страница 3: ...em is powered with a DC Power Supply through the IFD 5 Digital Interface Module as shown in Fig 1 0 The IFD 5 Module may be powered by any DC voltage between 12 and 24VDC The IFD 5 supplies power for...

Страница 4: ...vides a connector designed for use when the two modules are stacked When the two modules are stacked vertically the power input and digital communications connections are pin for pin directly above on...

Страница 5: ...dule is supplied configured for internal power To configure for external power a jumper inside the module must be moved The IFA 5 Module and the IFD 5 Module are similar in appearance so make sure you...

Страница 6: ...es above the graphical display Open the Pyrometer Settings menu for the pyrometer that is to be configured and follow the steps below 1 Select the Analog Output tab See Figure 5 below 2 Click the Use...

Страница 7: ...e table If the temperature goes below the temperature assigned to 1 volt the output will stay at 1 volt If the temperature exceeds the temperature assigned to 5 volts the output will stay at 5 volts T...

Страница 8: ...atchdog Failure Alarm Timeout In ____ box The Configure IFA 5 Module menu accessed through Menu Options Configure IFA 5 Module is used to set the Watchdog Failure Signal Level current or voltage level...

Страница 9: ...the meter should be connected in series 5 Configure the IFA 5 Module for Internal Power see instructions for Optional External Power 6 Make sure the IFA 5 Module has been running for at least 30 minu...

Страница 10: ...to 30VDC Power Consumption with IFD 5 IFA 5 and Pyrometer sensor 3 watts 125mA 24VDC Pin 1 10 Assignments Pin 1 GND Pins 1 3 5 and 10 are internally connected Pin 2 Isolates pyrometer if open COM pow...

Страница 11: ...racy 0 02 of FSR Resolution 0 002 of FSR Readback Accuracy 1 of FSR Zero Drift Voltage output 10 V C Current output 0 2 A C Span Temperature Coefficient 5ppm C Accuracy 0 02 of FSR Resolution 0 002 of...

Страница 12: ...and IFA 5 User Guide 9 5 1 IFD 5 and IFA 5 Support Cabinets Figure 8 Electronics support cabinet layout example Integrated Instrument System Figure 7 Electronics support cabinet layout example Fiber...

Страница 13: ...tput C Cooling Fan if used Active air cooling from the air purge system is the preferred method of cooling the cabinet as it reduces the risk of introducing contaminants to the cabinet D Digital Inter...

Страница 14: ...st temperature The temperature should not exceed 70 C Exceeding this temperature will not damage the electronics but may affect the temperature measurement accuracy If the temperature exceeds 80 C fai...

Отзывы: