BASF EXACTUS IFA-5 Скачать руководство пользователя страница 1

 

 

EXACTUS

®

 

IFD-5 DIGITAL and IFA-5 ANALOG OUTPUT 

User’s Guide 

V1.8  

 

BASF Corporation 

Temperature Sensing Group 

46820 Fremont Blvd. 

Fremont, CA  94538-6540 

Phone: (510) 490-2150 

Fax: (510) 252-1871 

http://www.catalysts.basf.com/tempsensing

 

Additional support available:  

BASF Temperature Sensing, Portland, Oregon 

4011 SE International Way, Suite 604 

Portland, OR 97222 

Phone: (503) 794-4073 

Fax: (503) 794-5591 

 

Copyright© 2014 BASF Corporation.  All rights reserved. 

 

Exactus is a registered trademark for which BASF Corporation has rights.  The data contained in this publication are based on our current 
knowledge and experience. In view of the many factors that may affect processing and application of our product, these data do not relieve 
processors from carrying out their own investigations and tests; neither do these data imply any guarantee of certain properties, nor the suitability 
of the product for a specific purpose. Any descriptions, drawings, photographs, data, proportions, weights, etc. given herein may change without 
prior information and do not constitute the agreed contractual quality of the product. It is the responsibility of the recipient of our products to 
ensure that any proprietary rights and existing laws and legislation are observed. 

Although all statements and information in this publication are believed to be accurate and reliable, they are presented gratis and for guidance 
only, and risks and liability for results obtained by use of the products or application of the suggestions described are assumed by the user. NO 
WARRANTIES OF ANY KIND, EITHER EXPRESS OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS 
FOR A PARTICULAR PURPOSE, ARE MADE REGARDING PRODUCTS DESCRIBED OR DESIGNS, DATA OR INFORMATION SET 
FORTH. Statements or suggestions concerning possible use of the products are made without representation or warranty that any such use is free 
of patent infringement and are not recommendations to infringe any patent. The user should not assume that toxicity data and safety measures are 
indicated or that other measures may not be required. ©2014 BASF Corporation. 

 

Содержание EXACTUS IFA-5

Страница 1: ...ta proportions weights etc given herein may change without prior information and do not constitute the agreed contractual quality of the product It is the responsibility of the recipient of our produc...

Страница 2: ...Optics either a lightpipe or lens assembly Ex 100 Electronics SST tube dimension 1 0 diameter 4 0 long 25 4mm 102mm IFD 5 Interface module for conditioning power and communicating with a PC Dimension...

Страница 3: ...em is powered with a DC Power Supply through the IFD 5 Digital Interface Module as shown in Fig 1 0 The IFD 5 Module may be powered by any DC voltage between 12 and 24VDC The IFD 5 supplies power for...

Страница 4: ...vides a connector designed for use when the two modules are stacked When the two modules are stacked vertically the power input and digital communications connections are pin for pin directly above on...

Страница 5: ...dule is supplied configured for internal power To configure for external power a jumper inside the module must be moved The IFA 5 Module and the IFD 5 Module are similar in appearance so make sure you...

Страница 6: ...es above the graphical display Open the Pyrometer Settings menu for the pyrometer that is to be configured and follow the steps below 1 Select the Analog Output tab See Figure 5 below 2 Click the Use...

Страница 7: ...e table If the temperature goes below the temperature assigned to 1 volt the output will stay at 1 volt If the temperature exceeds the temperature assigned to 5 volts the output will stay at 5 volts T...

Страница 8: ...atchdog Failure Alarm Timeout In ____ box The Configure IFA 5 Module menu accessed through Menu Options Configure IFA 5 Module is used to set the Watchdog Failure Signal Level current or voltage level...

Страница 9: ...the meter should be connected in series 5 Configure the IFA 5 Module for Internal Power see instructions for Optional External Power 6 Make sure the IFA 5 Module has been running for at least 30 minu...

Страница 10: ...to 30VDC Power Consumption with IFD 5 IFA 5 and Pyrometer sensor 3 watts 125mA 24VDC Pin 1 10 Assignments Pin 1 GND Pins 1 3 5 and 10 are internally connected Pin 2 Isolates pyrometer if open COM pow...

Страница 11: ...racy 0 02 of FSR Resolution 0 002 of FSR Readback Accuracy 1 of FSR Zero Drift Voltage output 10 V C Current output 0 2 A C Span Temperature Coefficient 5ppm C Accuracy 0 02 of FSR Resolution 0 002 of...

Страница 12: ...and IFA 5 User Guide 9 5 1 IFD 5 and IFA 5 Support Cabinets Figure 8 Electronics support cabinet layout example Integrated Instrument System Figure 7 Electronics support cabinet layout example Fiber...

Страница 13: ...tput C Cooling Fan if used Active air cooling from the air purge system is the preferred method of cooling the cabinet as it reduces the risk of introducing contaminants to the cabinet D Digital Inter...

Страница 14: ...st temperature The temperature should not exceed 70 C Exceeding this temperature will not damage the electronics but may affect the temperature measurement accuracy If the temperature exceeds 80 C fai...

Отзывы: