carried out by Bartels Mikrotechnik. Because of the mechanical fixation inside the electrical connector and due to 
the use of elevated driving voltages it is ensured that the pump performance is not affected by the surface corrosion. 
In a long term, we will work on improving the production process. Until then, referring to the unrestricted pump 
performance we ask our customers to excuse visible defects on the electrical connector. 

In general, the micropump can be driven with positive alternating voltages with maximum amplitude of 250 V 
at a frequency between 0 and 300 Hz.

 A rectangular signal results in best fluidic performance while a sine wave 

minimizes the audible noise. The actuators must be driven with a 180° phase shift in the signal. 

If a pump will be damaged while using a customer

’s controller, we do not provide any warranty.

 We recommend 

using our dedicated controllers. 

DANGER

 

THE MP6-SERIES MICROPUMPS ARE OPERATED AT HIGH VOLTAGES. BEFORE 

OPERATION, MAKE SURE THAT ALL SPECIFIC REGULATIONS FOR ELECTICAL SAFETY 

ARE FULLFILLED. 

Connecting the mp6-series to the cable 

The pumps can be connected via a FCC connector. The layout of the connector and pin assignment of the pumps is 
as shown below. Each piezo (P1 / P2) has a single lead for the negative (-P1/-P2) and the positive (+P1/+P2) supply 
voltage. 

The recommended connector  is a 4-pin 1.25 mm pitch FCC connector manufactured by Molex. The straight 
version part number 39 53 2045.

 An alternative is the angulated version with the part number 39 53 2044. 

The  connector  is  available  in  different  versions  to  fit  various  applications.  Reference  drawings  and  further 
specifications are available under www.molex.com 
The recommended maximum wire length between controller and the pump is 1 m. The customer has to assure that 
the assembly of pump, controller and electrical connection complies with the EMC regulations and electrical safety 
in the specific field of application. 
To connect the pumps and Molex, refer to following three figures. Orientate both components as indicated on the 
first picture, the pumps facing downwards with its serial number marking (!) and the Molex connector with the four 
small openings visible from above. Then insert the mp6/mp6-pp flex into the Molex connector (Step1).  Close the 
Molex connector to complete the interconnection between both components (Step 2). 

Содержание microComponents mp6-gas

Страница 1: ...1 Rev 3 1 08 2020 Operating Manual for Micropump mp6 series...

Страница 2: ...specifications 5 Technical Data of the mp6 hyb1 5 Technical Data of the mp6 liq1 2 6 Technical specifications mp6 pi1 7 Technical specifications mp6 pp1 9 Technical specifications mp6 gas1 10 Final in...

Страница 3: ...Low Voltage Directive 2014 35 EU Description of functions The micropumps have been developed for the transport of gases or liquids The controllers have been developed for operating the micropumps Bar...

Страница 4: ...ditions including vapors is not permitted Staff selection and qualification All work in connection with the installation assembly commissioning decommissioning disassembly operation servicing cleaning...

Страница 5: ...in typ flow rate p 0 8 ml min 4 typ back pressure p 0 500 mbar 7 25 psi 4 Gases air typ volume flow p 0 20 ml min 5 typ back pressure p 0 80 mbar 1 16 psi 5 1 Typical values Values can vary under appl...

Страница 6: ...Please contact us for more information Please find more information concerning the controller and the equipment in the corresponding manuals mp6 liq Order code mp6 liq Pump type piezoelectric diaphrag...

Страница 7: ...1 1811 x 0 5906 x 0 1498 in Weight 2 g Fluidic connectors barbed tube clip outer diameter 1 9 mm length 3 5 mm 2 Electric connector flex connector 1 25 mm pitch Power consumption 50 mW 5 Self priming...

Страница 8: ...Tygon tubing 1 3 mm inner diameter 3 Conditions Suction pressure 10 mbar DI water settings mp x 100 Hz 250 V SRS the max volume flow will be reached after a few minutes of operation time 4 Conditions...

Страница 9: ...ter Controllable flow range8 8 l min 4000 l min typ flow rate p 0 4 ml min 4 typ back pressure p 0 500 mbar 7 25 psi 4 1 Typical values Values can vary under application conditions Content is subject...

Страница 10: ...1498 in Weight 2 g Fluidic connectors barbed tube clip outer diameter 1 9 mm length 3 5 mm 2 Electric connector flex connector 1 25 mm pitch Power consumption 150 mW 5 Self priming yes 3 Pumping medi...

Страница 11: ...er 3 Conditions air room temperature 23 C settings mp x 300 Hz 250 V SRS the max volume flow will be reached after a few minutes of operation time 4 Conditions gases room temperature 23 C mp x 300 Hz...

Страница 12: ...rating the micropump In this chapter the operation of the micropump during the evaluation is described This chapter provides information on the proper connection of the pump with tubing and electrical...

Страница 13: ...mp6 series to the cable The pumps can be connected via a FCC connector The layout of the connector and pin assignment of the pumps is as shown below Each piezo P1 P2 has a single lead for the negative...

Страница 14: ...n top or bottom of the pump the Molex connector is not water tight additional sealing with e g silicone will be necessary fixation of the Molex connector in the final assembly is recommended The Molex...

Страница 15: ...neral for driving the pump with air the following points should be taken into account Higher frequencies should be used compared to pumping water Large amplitudes respectively driving voltages should...

Страница 16: ...uencies The performance depends on the environmental conditions Combination of micropumps The micropumps can be combined to achieve higher flow rates or pressure than the single unit can generate With...

Страница 17: ...l 49 231 47730 500 Fax 49 231 47730 501 Visit our Website www bartels mikrotechnik de for further information on applications Tutorials and helpful answers to frequently asked questions can be found i...

Отзывы: