27
Maintenance
Replacing the Lamp
1. Open the door on the right hand side of the
spectrophotometer. Push and twist counter-
clockwise to remove the lamp from the holder.
2. Insert the new lamp into the socket and care-
fully rotate clockwise until it locks into position.
3. Carefully wipe fingerprints off the lamp with a
small tissue.
4. Proceed to Realigning the Lamp.
Realigning the Lamp
It is necessary to realign the light bulb after each
transportation or replacement.
1. Turn on the power and the display should
show:
then press FUNC button and the display should
show:
2. Wait approximately one minute and the display
should show:
Caution
Unplug your spectrophotom-
eter before replacing the
lamp.
A
DJUST
L
AMP
L
OCATION
** CALIBRATION**
M
ODEL
-870
S
YSTEM
T
EST
P
ASS
0
A
DJUST
L
AMP
L
OCATION
26 ADC V
ALUE
:
XXXX
_
Caution
The lamp is hot when oper-
ating. Wait a few minutes
before touching the lamp.