![Barnstead International Thermolyne 859 Series Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/barnstead-international/thermolyne-859-series/thermolyne-859-series_service-manual_3942385080.webp)
80
Elapsed Time: 30:12
Sample Weight: 1093g
Weight Loss: 57.7g
Percent Loss: 5.27%
Temp Comp: 0.14%
Calib. Factor: 0.29%
===================
Calibrated Asphalt Ctnt
4.98%
===================
30 533 57.7 5.27
29 534 57.6 5.26
28 534 57.5 5.26
27 535 57.3 5.24
26 536 56.9 5.16
25 537 56.4 5.15
24 538 55.7 5.09
23 539 55.0 4.94
22 541 54.1 4.85
21 542 53.1 4.75
20 542 52.0 4.63
19 543 50.7 4.51
18 544 48.1 4.40
17 544 25.7 4.27
16 544 14.2 3.89
15 543 10.3 3.32
The sample did not lose 15 grams in the first 15 minutes.
14 542 9.9
2.21
This particular sample will run non-stop as a result!
13 540 9.2 2.01
12 536 8.9 1.99
11 534 8.4 1.54
10 512 8.1 1.02
9 511 6.2 0.95
8 492 5.2 0.81
7 481 5.1 0.76
6 477 3.2 0.45
5 472 2.1 0.24
4 469 0.9 0.05
3 465 0.7 0.04
2 459 0.2 0.02
1 451 0.1 0.01
-----:------:--------:------------
T: Temp: Wt. Loss: %Loss
----------------------------------
Filter Set Pt: 750°C
Chamber Set Pt: 538°C
Tested by:______________
Mix Type: ______________
Sample ID:______________
Time: 10:03:32
Date: 10-13-01
E
RRORS
Содержание Thermolyne 859 Series
Страница 1: ...NCAT Asphalt Content Furnace Service Manual LT1275X3 2 13 04...
Страница 4: ...4...
Страница 11: ...11 INSTALLATION...
Страница 12: ...12 INSTALLATION...
Страница 92: ...92...
Страница 93: ...93...
Страница 96: ...96...
Страница 97: ...97...
Страница 106: ...106...
Страница 107: ...107...
Страница 108: ...108...
Страница 109: ...109...