POL-200 Semiautomatic Polarimeter 

 

6. 

Place the standard tube into the sample chamber, press number keys to enter calibration value, press ENTER key to confirm. 

 

 

 
7.  

Press 

 or 

 key until visual field appear uniform brightness, press ENTER key to confirm. The instrument returns to zero-point automatically. 

8.    If you selected 3 points calibration, please repeat above steps until screen shows “Calibration is completed”. The instrument will return to 

 

measurement mode automatically. Calibration is completed. 

 

Restore Factory Settings 

If the instrument goes wrong during the use, you are able to follow the steps below to restore factory settings. NOTE: Reset function will restore the 
polarimeter back to factory default parameters, all calibration values and selected parameter will be lost or reset. 
1. 

Click the Setup icon to enter the setup menu. 

2. 

Click on the Reset icon, and press 

 or 

 key to select the “Enable” option. 

3. 

Press ENTER key, then press YES key. Reset is completed. 

 

 

Содержание POL-200

Страница 1: ...POL 200 Semiautomatic Polarimeter Instruction Manual BANTE INSTRUMENTS LIMITED...

Страница 2: ...ete If any are damaged or missing please contact nearest distributor Test tube 100 and 200 mm Sealing rings Power cable Power On Off Plug in the power supply and press ON OFF key to turn on the polari...

Страница 3: ...0 to 220 0mm Temperature Compensation Enable Default Disable Temperature compensation allows the instrument to run at ambient temperature but produces readings as if the system was running at 20 C Du...

Страница 4: ...ior to Use Rotate the focus knob on front panel until visual field became legible Zero Correction 1 Take a clean empty tube and place in the sample chamber 2 If the visual field appears the difference...

Страница 5: ...ke out the tube from the sample chamber Unscrew the cap remove the sealing ring and glass disc 2 Pour sample solution into the tube screw on the tube cap Wipe away the drops of water on glass disc 3 I...

Страница 6: ...red and recalled MEMORY INPUT 1 1 During the measurement click the Save icon the screen shows a dialog box and wait for entering the sample name 1 2 The default sample name is SAMPLE If you need to mo...

Страница 7: ...ring the measurement mode the screen shows the stored measured values 2 2 Press Page key the screen will toggle to next page 2 3 If you need to delete stored data press Delete key all of the data will...

Страница 8: ...p icon during the measurement mode the screen shows the setup menu 2 Click the Calibration icon the instrument enters the calibration mode 3 Follow the displayed guides to select the number of calibra...

Страница 9: ...ows Calibration is completed The instrument will return to measurement mode automatically Calibration is completed Restore Factory Settings If the instrument goes wrong during the use you are able to...

Страница 10: ...90 130 Z Accuracy 0 02 0 05 Z Resolution 0 005 Wavelength 589nm Light Source LED Interference filter Test Tube 50 100 200mm Temperature Range 0 100 C Temperature Accuracy 0 5 C Memory Stores up to 10...

Страница 11: ...w to speed up these efforts Warranty The warranty period for polarimeter is two years from the date of shipment Above warranty does not cover light source Out of warranty products will be repaired on...

Отзывы: