POL-200 Semiautomatic Polarimeter
8
6.
Place the standard tube into the sample chamber, press number keys to enter calibration value, press ENTER key to confirm.
7.
Press
or
key until visual field appear uniform brightness, press ENTER key to confirm. The instrument returns to zero-point automatically.
8. If you selected 3 points calibration, please repeat above steps until screen shows “Calibration is completed”. The instrument will return to
measurement mode automatically. Calibration is completed.
Restore Factory Settings
If the instrument goes wrong during the use, you are able to follow the steps below to restore factory settings. NOTE: Reset function will restore the
polarimeter back to factory default parameters, all calibration values and selected parameter will be lost or reset.
1.
Click the Setup icon to enter the setup menu.
2.
Click on the Reset icon, and press
or
key to select the “Enable” option.
3.
Press ENTER key, then press YES key. Reset is completed.