POL-200 Semiautomatic Polarimeter
3
SETTING THE PARAMETER:
1.
Click the Setup icon to enter the setup menu, select a parameter you want to set.
2.
Press
or
key to select an option or press the numeric keys to input the setting value.
3.
Press ENTER key to confirm. Setting is completed.
Prior to Use
Rotate the focus knob on front panel until visual field became legible.
Zero Correction
1.
Take a clean empty tube and place in the sample chamber.
2.
If the visual field appears the differences in apparent brightness, press ZERO key to enter the zero-point correction mode.
3.
Press
or
key to adjust the instrument until visual field appear the uniform brightness.
4.
Press ENTER key to confirm, the instrument returns to measurement mode. Zero correction is completed.