POL-200 Semiautomatic Polarimeter
7
Calibration
POL-200 semiautomatic polarimeter allows 1 to 3 points calibration, but calibration is not required unless the measured value appeared serious deviation.
1.
Click the Setup icon during the measurement mode, the screen shows the setup menu.
2.
Click the Calibration icon, the instrument enters the calibration mode.
3.
Follow the displayed guides to select the number of calibration points, press ENTER key confirm.
4.
Make sure that remove the tube from the sample chamber.
5.
Press
or
key until visual field appears uniform brightness, press ENTER key to confirm. Zero-point calibration is completed.