background image

POL-200 Semiautomatic Polarimeter 

 

Measurement 

POL-200 semiautomatic polarimeter can be used for measuring the four parameters, including the optical rotation, specific rotation, concentration and 
International Sugar Scale. Click the Setup icon to enter the setup menu, then click the “Measurement Mode” icon to select desired measurement mode.   
When the setting is completed, the instrument will automatically return to selected measurement mode. 
 
1. 

Take out the tube from the sample chamber. Unscrew the cap, remove the sealing ring and glass disc. 

 

 

 
2. 

Pour sample solution into the tube, screw on the tube cap. Wipe away the drops of water on glass disc.   

3. 

If necessary collect all air bubbles into the bubble trap. 

 

         

 

 
4. 

Place tube into the sample chamber. The bubble trap is upward.   

5. 

Press 

 or 

 key until visual field appears uniform brightness. Record the measured value. 

 

 

 
6. 

Press Return key, the instrument returns to the zero-point automatically. Measurement is completed. 

 

    Press and hold 

 or 

 key for 8 seconds, the instrument will enter auto-run mode. Press 

 or 

 key again, the instrument stops running. 

 
 
 

Bubble Trap 

Содержание POL-200

Страница 1: ...POL 200 Semiautomatic Polarimeter Instruction Manual BANTE INSTRUMENTS LIMITED...

Страница 2: ...ete If any are damaged or missing please contact nearest distributor Test tube 100 and 200 mm Sealing rings Power cable Power On Off Plug in the power supply and press ON OFF key to turn on the polari...

Страница 3: ...0 to 220 0mm Temperature Compensation Enable Default Disable Temperature compensation allows the instrument to run at ambient temperature but produces readings as if the system was running at 20 C Du...

Страница 4: ...ior to Use Rotate the focus knob on front panel until visual field became legible Zero Correction 1 Take a clean empty tube and place in the sample chamber 2 If the visual field appears the difference...

Страница 5: ...ke out the tube from the sample chamber Unscrew the cap remove the sealing ring and glass disc 2 Pour sample solution into the tube screw on the tube cap Wipe away the drops of water on glass disc 3 I...

Страница 6: ...red and recalled MEMORY INPUT 1 1 During the measurement click the Save icon the screen shows a dialog box and wait for entering the sample name 1 2 The default sample name is SAMPLE If you need to mo...

Страница 7: ...ring the measurement mode the screen shows the stored measured values 2 2 Press Page key the screen will toggle to next page 2 3 If you need to delete stored data press Delete key all of the data will...

Страница 8: ...p icon during the measurement mode the screen shows the setup menu 2 Click the Calibration icon the instrument enters the calibration mode 3 Follow the displayed guides to select the number of calibra...

Страница 9: ...ows Calibration is completed The instrument will return to measurement mode automatically Calibration is completed Restore Factory Settings If the instrument goes wrong during the use you are able to...

Страница 10: ...90 130 Z Accuracy 0 02 0 05 Z Resolution 0 005 Wavelength 589nm Light Source LED Interference filter Test Tube 50 100 200mm Temperature Range 0 100 C Temperature Accuracy 0 5 C Memory Stores up to 10...

Страница 11: ...w to speed up these efforts Warranty The warranty period for polarimeter is two years from the date of shipment Above warranty does not cover light source Out of warranty products will be repaired on...

Отзывы: