background image

POL-200 Semiautomatic Polarimeter 

 

Setup Menu 

POL-200 semiautomatic polarimeter contains an integrated setup menu that allows you to customize each option to meet measurement requirements. 
Selectable parameters contain the following options. 
 

 

 
INDEX: 

 

PARAMETER 

OPTION 

DESCRIPTION 

Measurement Mode 

    Optical Rotation 

    Specific Rotation 

    Concentration 

    International Sugar Scale 

Select your desired measurement mode. 

Tube length 

    100mm 

    200mm 

    Custom 

During the custom mode, the instrument acceptable setting ranges   
from 0.0 to 220.0mm. 

Temperature Compensation 

    Enable (Default) 

    Disable 

Temperature compensation allows the instrument to run at ambient   
temperature but produces readings as if the system was running at   
20ºC. During the Specific Rotation, Concentration and Sugar Degrees   
modes, the instrument will automatically enable the this feature. 

Concentration 

--- 

Sets the concentration of sample for measurement of specific rotation 

Specific Rotation 

--- 

Sets the specific rotation of sample for measurement of concentration 

Temperature Setting 

Range: 0.0 to 100.0

o

Calibrate the current temperature value 

Calibration 

1 to 3 points 

Using the standard tubes to calibrate the polarimeter 

Date and Time 

--- 

Sets the date and time for data log 

 

Содержание POL-200

Страница 1: ...POL 200 Semiautomatic Polarimeter Instruction Manual BANTE INSTRUMENTS LIMITED...

Страница 2: ...ete If any are damaged or missing please contact nearest distributor Test tube 100 and 200 mm Sealing rings Power cable Power On Off Plug in the power supply and press ON OFF key to turn on the polari...

Страница 3: ...0 to 220 0mm Temperature Compensation Enable Default Disable Temperature compensation allows the instrument to run at ambient temperature but produces readings as if the system was running at 20 C Du...

Страница 4: ...ior to Use Rotate the focus knob on front panel until visual field became legible Zero Correction 1 Take a clean empty tube and place in the sample chamber 2 If the visual field appears the difference...

Страница 5: ...ke out the tube from the sample chamber Unscrew the cap remove the sealing ring and glass disc 2 Pour sample solution into the tube screw on the tube cap Wipe away the drops of water on glass disc 3 I...

Страница 6: ...red and recalled MEMORY INPUT 1 1 During the measurement click the Save icon the screen shows a dialog box and wait for entering the sample name 1 2 The default sample name is SAMPLE If you need to mo...

Страница 7: ...ring the measurement mode the screen shows the stored measured values 2 2 Press Page key the screen will toggle to next page 2 3 If you need to delete stored data press Delete key all of the data will...

Страница 8: ...p icon during the measurement mode the screen shows the setup menu 2 Click the Calibration icon the instrument enters the calibration mode 3 Follow the displayed guides to select the number of calibra...

Страница 9: ...ows Calibration is completed The instrument will return to measurement mode automatically Calibration is completed Restore Factory Settings If the instrument goes wrong during the use you are able to...

Страница 10: ...90 130 Z Accuracy 0 02 0 05 Z Resolution 0 005 Wavelength 589nm Light Source LED Interference filter Test Tube 50 100 200mm Temperature Range 0 100 C Temperature Accuracy 0 5 C Memory Stores up to 10...

Страница 11: ...w to speed up these efforts Warranty The warranty period for polarimeter is two years from the date of shipment Above warranty does not cover light source Out of warranty products will be repaired on...

Отзывы: