POL-200 Semiautomatic Polarimeter
2
Setup Menu
POL-200 semiautomatic polarimeter contains an integrated setup menu that allows you to customize each option to meet measurement requirements.
Selectable parameters contain the following options.
INDEX:
PARAMETER
OPTION
DESCRIPTION
Measurement Mode
●
Optical Rotation
●
Specific Rotation
●
Concentration
●
International Sugar Scale
Select your desired measurement mode.
Tube length
●
100mm
●
200mm
●
Custom
During the custom mode, the instrument acceptable setting ranges
from 0.0 to 220.0mm.
Temperature Compensation
●
Enable (Default)
●
Disable
Temperature compensation allows the instrument to run at ambient
temperature but produces readings as if the system was running at
20ºC. During the Specific Rotation, Concentration and Sugar Degrees
modes, the instrument will automatically enable the this feature.
Concentration
---
Sets the concentration of sample for measurement of specific rotation
Specific Rotation
---
Sets the specific rotation of sample for measurement of concentration
Temperature Setting
Range: 0.0 to 100.0
o
C
Calibrate the current temperature value
Calibration
1 to 3 points
Using the standard tubes to calibrate the polarimeter
Date and Time
---
Sets the date and time for data log