BANTE320 PORTABLE PH/ION METER
27
Optional Ion Selective Electrodes
The meter can connect with various types of composite ion selective electrodes. The following list shows the selectable sensors.
MODEL
ION
F-US
Fluoride
CL-US
Chloride
Br-US
Bromide
Cn-US
Cyanide
Na-US
Sodium
NO3-US
Nitrate
Ca-US
Calcium
NH4-US
Ammonium
Cd-US
Cadmium
NH3-US
Ammonia
Cu-US
Cupric
I-US
Iodide
Pb-US
Lead
K-US
Potassium
Ag-US
Silver
S-US
Sulphide
Troubleshooting
LCD DISPLAY
CAUSE
CORRECTIVE ACTION
Electrode dried out
■
Soak the pH electrode in 3M KCL solution for 10 minutes
■
Soak the Ion Selective Electrode in the diluted standard
solution at least 15 minutes
Measured value is out of range
■
Check the electrode whether clogged, dirty or broken
Incorrect calibration solutions
■
Use fresh calibration solutions for calibration
Electrode is out of service life
■
Replace the pH or Ion Selective Electrode
Keypad is not working properly
■
Replace the batteries