Appendix

 

 

Appendix   

 

Indexing a sample vial 

The United States Environmental Protection Agency (U.S. EPA) recommends that the vial used for turbidity calibration or sample measurement be 
indexed. Its purpose is obtain a position which provides the lowest turbidity reading. The indexing methods are as follows. 
1.1  Fill the vial using the distilled water (< 0.5 NTU). Cap the vial. 
1.2  Wipe the vial with the lint-free cloth to remove waterdrops and fingerprints. 
1.3  Place the vial in the measurement chamber. Align the mark on the vial with the arrow on the meter. 
1.4  Press the 

Meas

 key, the meter begins the measurement. 

1.5  Slowly rotate the vial approximately 45 degrees. Close the light shield lid and record the measured value. 
1.6  Repeat the step 1.5 until the lowest turbidity reading is shown. 
1.7  Mark this position on the vial. 
 

Matching the sample vials 

For the best accuracy and repeatability, using one indexed vial is best choice for turbidity measurement. If you need to use a few vials, match   
these vials are necessary. 
2.1  Repeat the steps above for each vial and record the measured values. 
2.2  Find the closest position of these vials measuring value and mark it. 
 

Preparing the formazin standards 

Preparation of turbidity-free water: 
The turbidity-free water is used for preparation of turbidity standards and is prepared by filtering deionized water through a 0.45µ m or smaller 
pore-sized membrane. 
 
Preparation of turbidity standards: 

 

Turbidity Standard 

Preparation Method 

4000 NTU 

 

Dissolve 1g hydrazine sulfate [(NH

2

)

2

H

2

SO

4

] in the turbidity-free water and dilute to 100ml in a volumetric flask. 

 

Dissolve 10g hexamethylenetetramine [(CH

2

)

 

6

 

N

4

] in the turbidity-free water and dilute to 100ml in a volumetric

.

 

 

flask. 

 

Mix 5ml of hydrazine sulfate and 5ml of hexamethylenetetramine solutions in a 100ml volumetric flask and let

.

   

 

stand 24 hours at 25

 

(± 3)° C. 

1000 NTU 

 

Mix 25ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

500 NTU 

 

Mix 12.5ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

200 NTU 

 

Mix 10ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

 

Mix 50ml of above standard in a 100ml volumetric flask and dilute to the mark. 

10 NTU 

 

Mix 10ml of 4000 NTU standard in a 100ml volumetric flask and dilute to the mark. 

 

Mix 2.5ml of above standard in a 100ml volumetric flask and dilute to the mark. 

 

Or,

 

mix 2ml of 500

 

NTU standard in a

 

100ml volumetric flask and dilute to the mark. 

 

 

 

 

 

 

Содержание TB100

Страница 1: ...TB100 Portable Turbidity Meter Instruction Manual Bante Instruments Inc...

Страница 2: ...ber of calibration points 7 Calibrating the meter 8 Turbidity measurement 8 Selecting the measurement mode 8 Single measurement low turbidity samples 200 NTU 9 Single measurement high turbidity sample...

Страница 3: ...of meter After the unpacking please check all components are complete If any are damaged or missing please contact the supplier The sample vial in the measurement chamber has been indexed with the 0...

Страница 4: ...ment mode Starts calibration Enters the setup menu Press and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the data log Decrease v...

Страница 5: ...meter is in the calibration mode Calibration icon Indicates the meter is calibrating Installing the Batteries 1 Remove the battery compartment cover from the backside of meter 2 Insert the three AA ba...

Страница 6: ...fault turbidity unit NTU Nephelometric turbidity unit NTU FNU Formazin nephelometric unit EBC Turbidity unit of the European Brewery Commission ASBC Turbidity unit of the American Society of Brewing C...

Страница 7: ...e factory settings The Factory Reset option is used to restore the meter back to default settings If enabled all of the calibration data and selected options will be lost or reset the meter must be re...

Страница 8: ...eaned thoroughly with deionised water after each measurement Any residues can cause inaccurate readings Turbidity Calibration The TB100 portable turbidity meter is calibrated with Formazin Standards a...

Страница 9: ...ext calibration point 2 7 If necessary press the key to select an applicable calibration point 2 8 Repeat steps 2 2 through 2 5 until the meter shows Calibration is completed To exit the calibration m...

Страница 10: ...flashing 1 6 Wait for the reading to stabilize the Measure icon will stop flashing Press the Meas key the meter will take a measurement again Single measurement high turbidity samples 1100 NTU For the...

Страница 11: ...e key the display shows the date and time of measurement Format mm dd hh mm 3 Press the key the display shows the stored data 4 Press the key again the display shows the next data set 5 Press the Meas...

Страница 12: ...computer 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the d...

Страница 13: ...f these vials measuring value and mark it Preparing the formazin standards Preparation of turbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared b...

Страница 14: ...Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 100 data sets Output USB communication interface Display Segmented LCD 60 L 40 W mm Operating Temperatu...

Страница 15: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Отзывы: