Setup Menu

 

 

Switching the Meter On and Off 

•   

  Press and release the

 

 

key to switch on the meter. 

•   

  Press and hold the

 

 

key for 3 seconds to switch off the meter. 

 

 

   

 

 

Setup Menu 

The TB100 portable turbidity meter contains an integrated setup menu that is used to customize the displayed option to meet measurement   
requirement. The following table describes the functions of each menu item. 

 

Menu 

Description 

Options 

Description 

Default 

 

Set the number of calibration points 

 

2 to 5 points 

 

Set the date and time 

--- 

Year, month, day, hour, minute 

--- 

 

Set the default turbidity unit 

NTU 

Nephelometric turbidity unit 

NTU 

FNU 

Formazin nephelometric unit 

EBC 

Turbidity unit of the European Brewery Commission 

ASBC 

Turbidity unit of the American Society of Brewing Chemists 

 

Set the resolution of measurement 

 

0.01 

0.01 

 

0.1 

 

Set the measurement mode 

 

Single measurement 

Yes 

 

Continuous measurements 

 

Auto-power off 

 

Enable 

No 

 

Disable 

 

Clear all data logs 

 

Enable 

No 

 

Disable 

 

Reset the meter to default settings 

 

Enable 

No 

 

Disable 

 

Setting the default option 

1.1  Press and hold the

 

 

key for 3 seconds to enter the setup menu. 

1.2  Press the 

 / 

 key to select a menu item, press the 

Enter

 key to confirm. 

1.3  Press the 

 / 

 key to select a desired option, press the 

Enter

 key to confirm and return to the measurement mode. 

3 sec. 

Содержание TB100

Страница 1: ...TB100 Portable Turbidity Meter Instruction Manual Bante Instruments Inc...

Страница 2: ...ber of calibration points 7 Calibrating the meter 8 Turbidity measurement 8 Selecting the measurement mode 8 Single measurement low turbidity samples 200 NTU 9 Single measurement high turbidity sample...

Страница 3: ...of meter After the unpacking please check all components are complete If any are damaged or missing please contact the supplier The sample vial in the measurement chamber has been indexed with the 0...

Страница 4: ...ment mode Starts calibration Enters the setup menu Press and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the data log Decrease v...

Страница 5: ...meter is in the calibration mode Calibration icon Indicates the meter is calibrating Installing the Batteries 1 Remove the battery compartment cover from the backside of meter 2 Insert the three AA ba...

Страница 6: ...fault turbidity unit NTU Nephelometric turbidity unit NTU FNU Formazin nephelometric unit EBC Turbidity unit of the European Brewery Commission ASBC Turbidity unit of the American Society of Brewing C...

Страница 7: ...e factory settings The Factory Reset option is used to restore the meter back to default settings If enabled all of the calibration data and selected options will be lost or reset the meter must be re...

Страница 8: ...eaned thoroughly with deionised water after each measurement Any residues can cause inaccurate readings Turbidity Calibration The TB100 portable turbidity meter is calibrated with Formazin Standards a...

Страница 9: ...ext calibration point 2 7 If necessary press the key to select an applicable calibration point 2 8 Repeat steps 2 2 through 2 5 until the meter shows Calibration is completed To exit the calibration m...

Страница 10: ...flashing 1 6 Wait for the reading to stabilize the Measure icon will stop flashing Press the Meas key the meter will take a measurement again Single measurement high turbidity samples 1100 NTU For the...

Страница 11: ...e key the display shows the date and time of measurement Format mm dd hh mm 3 Press the key the display shows the stored data 4 Press the key again the display shows the next data set 5 Press the Meas...

Страница 12: ...computer 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the d...

Страница 13: ...f these vials measuring value and mark it Preparing the formazin standards Preparation of turbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared b...

Страница 14: ...Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 100 data sets Output USB communication interface Display Segmented LCD 60 L 40 W mm Operating Temperatu...

Страница 15: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Отзывы: