Communication

 

 

Communication 

The TB100 portable turbidity meter is capable of transfering the data to computer or importing the data to Excel by a DAS software. You are able 
to download this software from our official website. Before installation, ensure that Windows 7/8/10 operating system has been installed on   
your computer. 
 

Receiving the data 

1. 

Connect the USB cable to meter and computer. Click the 

DAS

 icon, the system will automatically scan an available communication port 

 

and show the message box “Found a port on your computer”.

 

2. 

Click the 

OK

 button, the application starts. 

 

  If your computer can not find a communication port, please click the file 

CP210xVCPInstaller_x64.exe

 or 

CP210xVCPInstaller_x86.exe

 

 

to update the drive program. 

 

 

 
3. 

Click the 

Connect

 

button, the screen shows “Port is connected” indicat

ing that the communication between the meter and the computer 

 

has been established.   

4. 

Click the 

OK

 button to confirm. 

5. 

Click the 

Receive

 button, the stored data automatically transfer to computer. 

 

Create the Excel file 

When the transfer is completed, click the 

Save as Excel

, the measured values in the data sheet will automatically convert to Excel file.   

 

  WARNING: once the software is closed, all received data will be lost and can not be recovered. 

 
 
 
 
 
 
 
 
 
 
 
 

Содержание TB100

Страница 1: ...TB100 Portable Turbidity Meter Instruction Manual Bante Instruments Inc...

Страница 2: ...ber of calibration points 7 Calibrating the meter 8 Turbidity measurement 8 Selecting the measurement mode 8 Single measurement low turbidity samples 200 NTU 9 Single measurement high turbidity sample...

Страница 3: ...of meter After the unpacking please check all components are complete If any are damaged or missing please contact the supplier The sample vial in the measurement chamber has been indexed with the 0...

Страница 4: ...ment mode Starts calibration Enters the setup menu Press and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the data log Decrease v...

Страница 5: ...meter is in the calibration mode Calibration icon Indicates the meter is calibrating Installing the Batteries 1 Remove the battery compartment cover from the backside of meter 2 Insert the three AA ba...

Страница 6: ...fault turbidity unit NTU Nephelometric turbidity unit NTU FNU Formazin nephelometric unit EBC Turbidity unit of the European Brewery Commission ASBC Turbidity unit of the American Society of Brewing C...

Страница 7: ...e factory settings The Factory Reset option is used to restore the meter back to default settings If enabled all of the calibration data and selected options will be lost or reset the meter must be re...

Страница 8: ...eaned thoroughly with deionised water after each measurement Any residues can cause inaccurate readings Turbidity Calibration The TB100 portable turbidity meter is calibrated with Formazin Standards a...

Страница 9: ...ext calibration point 2 7 If necessary press the key to select an applicable calibration point 2 8 Repeat steps 2 2 through 2 5 until the meter shows Calibration is completed To exit the calibration m...

Страница 10: ...flashing 1 6 Wait for the reading to stabilize the Measure icon will stop flashing Press the Meas key the meter will take a measurement again Single measurement high turbidity samples 1100 NTU For the...

Страница 11: ...e key the display shows the date and time of measurement Format mm dd hh mm 3 Press the key the display shows the stored data 4 Press the key again the display shows the next data set 5 Press the Meas...

Страница 12: ...computer 2 Click the OK button the application starts If your computer can not find a communication port please click the file CP210xVCPInstaller_x64 exe or CP210xVCPInstaller_x86 exe to update the d...

Страница 13: ...f these vials measuring value and mark it Preparing the formazin standards Preparation of turbidity free water The turbidity free water is used for preparation of turbidity standards and is prepared b...

Страница 14: ...Detector Silicon photodiode Stray Light 0 02 NTU Sample Vial 60 H 25 Dia mm Memory Stores up to 100 data sets Output USB communication interface Display Segmented LCD 60 L 40 W mm Operating Temperatu...

Страница 15: ...r suggestions on how to speed up these efforts Warranty The warranty period for meter is two years from the date of shipment Above warranty does not cover the calibration standards Out of warranty pro...

Отзывы: