6536-000 en/2022-03
33/54
Operation
/5
Temperature of the sonication
fluid below 20 °C
• Replace all or part of the
sonication fluid
• Degas the sonication fluid
again, see
• Abort cancels the pro-
cess. After cancelling, the
startup screen is dis-
played.
• OK resumes the process.
The robotics program was
started, but a TRISON Twist
was not connected
• Attach TRISON Twist, see
• Abort cancels the pro-
cess. After cancelling, the
startup screen is dis-
played.
• OK resumes the process
without moving function.
Log memory full
• Store and delete logs, see
displays the "Documenta-
tion" screen.
• Cancel closes the mes-
sage. Logs are not saved.
After cancelling, the
startup screen is dis-
played.
Maintenance required by the
manufacturer
• Contact the manufacturer,
see
. OK dis-
plays the "Information"
screen with contact data
and information on the
ultrasonic bath.
• Cancel closes the mes-
sage. After cancelling, the
startup screen is dis-
played.
• Filter clogged or installed
incorrectly
• Pressure sensor defective
• Replace or rinse filter,
check installation, see
chapter
• Contact the manufacturer,
see
Warning screen
Reasons
Measures
Содержание TRISON