Design and function
/3
3 Comb bar for 8 adapters
4 Support for the oscillating tank
The TRISON Rack is used to clean up to 8 rinsable MIS instruments simultane-
ously. The TRISON Rack is available as a left- or right-hand variant, see chapter
The instruments are connected to the corresponding adapters, rinsed from the
inside and individually checked for flow. Instruments that are not rinsable are dis-
played on the touchscreen after cleaning.
3.7 Temperature sensor
Fig. 8
Temperature sensor (optional)
1 Fastening clamp
2 Immersion sensor
3 Plug for connection to the TRISON Base
4 LED display
Information
The plug on the temperature sensor may not be immersed in liquid.
– If the plug has come into contact with liquid, allow it to dry completely before
connecting it.
– The plug can be blown out with compressed air.
The temperature sensor is used to monitor the temperature during cleaning of
standard instruments.
Meaning of the LED displays:
– Green, steady light: Bath temperature OK
– Red, steady light: Bath temperature too low
– Red, flashing: Bath temperature too high
– Orange, flashing: Immersion sensor not immersed