Ø 40 mm
Ø 60 mm
Ø 7 mm
(in base
M6
thread /
length
max. 8 mm
)
14 / 25
61279e GB/2018-03
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
• From a production perspective, the layout of the
fi
xation holes cannot be speci
fi
ed
beforehand. Therefore
- Place the control unit on its side and turn the base through 6 lock-in positions
(clicks), starting from the left or right limit position (= midpoint of the subsequent
swivel range).
- Mark the drill holes on the outer edge of the base (can be wiped off).
- Position the control unit on its future
fi
xation spot in accordance with the position
diagram on the next page, and turn it toward the user.
- Then, transfer the marked drill holes on the base to the work surface.
• Use and position the
fl
at seal as a template, drill and deburr the drill holes.
• Align the
fl
at seal toward the work surface, in accordance with the drill holes.
• Screw in the threaded rods as mounting aids in the base.
• Guide both connectors through the work surface.
• Align the control unit while inserting the threaded rods through the corresponding
drill holes.
• Now secure the base from the bottom onto the remaining holes using screws.
Remove the threaded rods and replace them with 2 screws.
• Set up the compressed air connection under the table top. To do so, use the coupling
socket supplied. See chapter 8.
• If desired, connect the Ethernet cable to the corresponding interface of the control
unit.
IMPORTANT:
The speci
fi
ed degree of protection for the control unit is only preserved by using the
supplied Ethernet connector.
To do this, mount the plug to an existing, conventional Ethernet cable.
Assembly diagrams
The following drill holes must be
provided in the work surface for
assembly purposes.