Warning and safety precautions
• Installation to be conducted
only
by authorised skilled personnel!
• TRISON ultrasonic baths must be installed in accordance with these installation
instructions.
If you have any questions, please contact the manufacturer.
• Do not weld the stainless steel ultrasonic oscillating tank to the work surface!
• Inspect all the individual parts for possible transportation damage after removing
them from the packaging!
If transportation damage is found, do not connect the TRISON ultrasonic bath to
the mains. Instead, report the damage immediately and in writing to the delivering
shipping agent and the supplier! Keep the original packaging!
•
Only
plug in the ultrasonic bath to an outlet with a
grounded socket
!
• Control units (TB ..., ST ...), HF generators and mains supply switches may only be
opened by authorised skilled personnel.
• Defective parts may only be replaced with original parts!
• Never operate the ultrasonic oscillating tank without the cover on the transducers!
• Keep the HF generator and operating elements
clean and dry
!
• Only operate the mains supply switch from the ST ... with a cover plate or plugged-in
conductor!
• Do
not
operate the ultrasonic oscillating tank without
fl
uids!
• Follow the instructions for use!
• Cavitation noises are emitted during ultrasonic operation.
When working continuously within a radius of 2 m from the ultrasonic oscillating tank,
wear hearing protection to protect against injury!
• Do
not
place anything directly on the bottom of the tank; place the goods to be
treated in the appropriate accessories (TRISON Twist, TRISON Rack)
• Position/suspend the HF generator and mains supply switch on a dry and
fi
rm
surface.
• Please note that the ultrasonic oscillating tank will be delivered with short-circuited
oscillation systems.
Before starting up the short-circuit bridge at the ultrasonic oscillating tank or at the HF
connection has to be removed.
• Only turn on the HF generator if the ultrasonic oscillating tank is connected (HF
cable) – never run the HF generator without a load.
• Do not operate the ultrasonic bath while unattended.
• Do
not
fi
ll the ultrasonic oscillating tank with
hot
water.
The
fi
lling temperature may not be higher than 50°C.
• All live conductors must be protected from liquids – chapter 8.