6486-001 GB/2021-04
25/36
Maintenance
/6
7. The foil should be perforated, see illustration. Oth-
erwise, we recommend having the device checked
by the BANDELIN electronic GmbH & Co. KG service
department: see chapter
6.4 Repair
.
8. Archive the foil with the test date and the serial
number of the ultrasonic bath. In addition, the doc-
umentation template for the foil test can be filled in
and archived.
9. Rinse the ultrasonic oscillating tank thoroughly to remove loose foil particles.
You can order suitable foil test frames from BANDELIN electronic GmbH & Co. KG.
The foil test frames are configured for a wide range of tank dimensions. Alumin-
ium foil is additionally required to perform the test as it is not included in the
delivery.
Type
Order no. For
FT 36
3673
ZE 1031/1032/... DT
FT 37
3674
ZE 1058/1059/... DT