5
/
Operation
16/36
6486-001 GB/2021-04
5.3 Sonication time
NOTICE
Risk of damage to the goods to be treated
Excessively long sonication can damage the surface of the goods to be treated.
– Select a short sonication time where possible.
The ideal sonication time depends on several factors:
• Type and concentration of agent
• Working temperature of sonication fluid
• Type of soiling
• Type of goods to be treated, in particular materials
Note the information on the recommended sonication time from the agent man-
ufacturer. To begin with, select as short a sonication time as possible to protect
the objects to be sonicated and the ultrasonic oscillating tank. Check the result.
Lengthen the sonication time if the result is not sufficient.
5.4 Pour in sonication fluid
CAUTION
Risk of scalding
– Do not pour hot water into the ultrasonic oscillating tank.
– Maximum filling temperature: 50°C.
NOTICE
Damage due to condensation in the ultrasonic bath
At high humidity, condensation forms inside the device if you fill it with cold
water.
– Do not pour cold water into the ultrasonic oscillating tank when humidity is
high.
10 l x 2,5 %
= 0,25 l agent
100 %
10 l - 0,25 l = 9,75 l water